Calibration: Autocal/Drift Values; Calibration: Autocal/Drift Values: Drift Values - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

8.2.7

Calibration: AUTOCAL/drift values

Thermo AUTOCAL: OFF
Drift values
Cycle time AUTOCAL
Purge time
8.2.7.1

Calibration: AUTOCAL/drift values: Drift values

This function is used to display the drift values of the zero point and sensitivity, and to change
them if necessary. The drift values are the total of the deviations in measured values for the zero
and sensitivity calibrations. This parameter is specific to the component.
The drift parameters of the UV measuring ranges differ from those of the IR measuring ranges.
Select component
: SO2
Drift
Zero:
MR1:
Reset drift values
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
It is possible to change the following parameters in this dialog:
• Thermo-AUTOCAL
827
• Drift values
• Cycle time
• Purge time
This menu display appears following selection of the drift values
function and allows selection of the desired component.
1
You can switch between the individual components by pressing
an arrow key. You can select the displayed component using the
8271
<ENTER> key.
You can now view the drift values and reset them if necessary. To
SO2
515 vpm
do this, position the cursor on the 4th line (reset) of the display
0
0.0
MR2:
0.0
and press the <ENTER> key.
82711
– OFF: An automatic AUTOCAL only takes place when the
set cycle time has expired (see there).
– ON: An automatic AUTOCAL only takes place when the
cycle time has expired. In addition, an AUTOCAL is trig‐
gered if the operating temperature has changed by more
than 8 °C compared to the temperature measured during
the last AUTOCAL. This AUTOCAL it started with a delay of
280 minutes.
These three parameters are described separate‐
ly.
Functions
8.2 Calibration
125