Siemens ULTRAMAT 23 Series Equipment Manual page 200

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Spare parts/accessories
13.6 IR analyzer units
Part No.
Designation
1 *)
IR source with temperature sensor
2.1
Spacer
2.2 **)
Optical filter
3 *)
Chopper 1-channel or 2-channel
4
Chopper plate glued 1-channel or 2-
channel
5.1
O-Ring for analyzer chamber
5.2
Analyzer chamber
5.3
Analyzer chamber
6
Gas filter
7.1/7.2 *
Receiver chamber
**)
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certified service center
or by qualified, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.
200
Remarks
For IR source, if optical filter available
Depends on selected measuring range and measured component
Version depending on the configuration
Version depending on the configuration with threaded rods and win‐
dows
Required for analyzer chamber lengths 20 mm and more
Without o-Ring
Depends on selected measuring range and measured component
Without o-Ring
Depends on selected measuring range and measured component
Depends on selected measuring range and measured component
Depends on the selected measured component
Equipment Manual, 04/2022, A5E37100388-012