Analyzer Status: Status: O2 Sensor Status; Analyzer Status: Status: H2S Sensor Status; Diagnostics: Diagnostics Values - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

8.1.1.4

Analyzer status: Status: O2 sensor status

O2 sensor status
Reading:
Maint. req: 6.00 mV
Fault:
8.1.1.5

Analyzer status: Status: H2S sensor status

H2S sensor status
Reading:
Maint.req :373.50 nA
Fault:
8.1.2

Diagnostics: Diagnostics values

The diagnostic values provide important information for troubleshooting and adjustments.
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
The probe voltage of the O
its aging process. Therefore the probe voltage is measured with
11.11 mV
each AUTOCAL. A warning (maintenance request) is output if
5.00 mV
the value drops below 6.0 mV. The oxygen sensor should there‐
fore be replaced when this value is reached. An exact measure‐
ment is no longer possible if the probe voltage falls below the
minimum value of 5.0 mV (fault message "Sensitivity of O
sor too low").
• The reading (actual value) is the probe voltage measured on
the sensor during the last AUTOCAL.
• The warning (maintenance request) and fault are the two
minimum values where a maintenance request or fault mes‐
sage is output when fallen below.
The H
2
steadily decreasing its sensitivity. If a value below the minimum
747.00 nA
value for a warning is determined during calibration of the sen‐
298.80 nA
sitivity, the service life of the sensor has almost been reached
8115
(maintenance request). If the sensitivity drops further below the
value for a fault, the fault message "Sensitivity of H
low" is output. The sensor must then be replaced.
• The reading (actual value) is the sensor sensitivity measured
during the last sensitivity calibration.
• The warning (maintenance request) and fault are the two
minimum values where a maintenance request or fault mes‐
sage is output when fallen below.
sensor is reduced during use due to
2
S sensor ages with increasing operating time, thereby
Functions
8.1 Diagnostics
sen‐
2
S sensor too
2
105