Siemens ULTRAMAT 23 Series Equipment Manual page 86

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Commissioning
6.4 System setup with several analyzers in parallel
Figure 6-2
Parallel connection without internal pump and solenoid valve switching
Parameter assignments
Master:
• Enter the AUTOCAL cycle time, e.g.: 6 hours (see Calibration: AUTOCAL/drift values: Cycle
time (Page 126)).
• Assign the "Sync." function to a relay (see Configuration: Inputs/outputs/pump: Assign relays
(Page 140)).
• Assign the "Zero gas" function to a relay (see Configuration: Inputs/outputs/pump: Assign
relays (Page 140)).
• Assign the "Only CAL contact" function to the digital input SYNC (see Configuration: Inputs/
outputs/pump: Binary/sync inputs (Page 143)).
Slave:
• Set the AUTOCAL cycle time to "0" to prevent a cyclic AUTOCAL from being triggered
(see Calibration: AUTOCAL/drift values: Cycle time (Page 126)).
• Assign the "Sync." function to a relay (see Configuration: Inputs/outputs/pump: Assign relays
(Page 140)).
• Assign the "AUTOCAL" function to the digital input SYNC (see Configuration: Inputs/outputs/
pump: Binary/sync inputs (Page 143)).
86
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23