5.3 Alarm functions

Instrument alarm
60
Remote alarm
Leica IP S
The Leica IP S is equipped with two different alarm functions:
The printer has a beeper that emits acoustic signals indicating important in-
strument states and functions.
Button pressed:
Magazine empty/tray stack full: 2 short beeps
Error:
Back of instrument
56
This alarm is external to the Leica IP S.
• The remote alarm device is connected to the printer via a 3.5 mm-diameter
jack connector (optional accessory), which is connected to socket (56) .
• The remote alarm is generated in the instrument when there is a general
power failure (no mains power) or when the printer is switched off with
main switch (53) at the rear.
The remote alarm device connected to the instrument must be
rated at less than 100 mA. A maximum voltage of 24 V DC may
be present.
For details on how to connect a remote alarm device to the
Leica Printer IP C, please contact your local Leica sales office
or the manufacturer Leica Microsystems Nussloch GmbH di-
rectly.
1short beep
5 short beeps
The beeper can be deactivated
by means of the DIL- switches at
53
the back panel of the printer.
To deactivate the beeper, push
the bottommost switch (60) to the
right (in the direction of the red
arrow in
Fig. 29
5. Operation
Fig.
29).