Leica IP S Operating Instructions Manual page 27

Automated printing system for microscope slides
Table of Contents

Advertisement

87
Leica IP S
86
Fig. 24
• Insert the cartridge into the receptacle located
behind the cover on the left side of the printer.
Loosen the red screw-on cap one full turn but
do not yet remove it.
• Insert the cartridge completely into the recep-
tacle applying moderate pressure. You have
to use a certain amount of force to pierce the
seals.
• Withdraw the cartridge by about 1½ inch (30
mm) and push it back in as far as it will go,
repeat this step three times.
• Unscrew the red cap completely from the
nozzle of the cartridge and store the screw-
on cap in the groove indented into the car-
tridge for that purpose.
Important!
Prior to each transport, the cap must
be screwed onto the nozzle to pre-
vent the ink from spilling.
• Directly next to the cartridge receptacle is a
safety catch. To ensure safe operation of the
printer, slide the safety catch sidewise in front
of the cartridge to prevent the cartridge from
sliding out of the receptacle.
This step is very important to ensure
safe operation of the printer!
• If your particular instrument is not equipped
witha safety catch, look for it in the accesso-
ries kit.
Before installing the cartridge, you need to
install the safety catch, in unbolted position.
• If there is no safety catch in your printer nor is
it included in the accessories kit, please con-
tact your sales representative.
4. Setup
27