SYSTem . . . :TERMinator:TIMEout < timeout >
SYSTem . . . :TERMinator:TIMEout?
SYSTem . . . :TRANsmit:AUTO 1 | 0 | ON | OFF
SYSTem . . . :TRANsmit:AUTO?
SYSTem . . . :TRANSmit:BAUD < baud_rate >
SYSTem . . . :TRANSmit:BAUD?
This command sets the termination timeout. If one
SENSe:SERial[1|2|3|4]:DATA?
period, the data that was received is sent to the host computer. If one
SOURce:SERial[1|2|3|4]:TEXT?
cannot be completed within the timeout period, the command is aborted and a
transmit error is added to the error queue. <time_out> can be set from 0 to 65535 (in
seconds). If <time_out> is 0, the timeout function is disabled. Power-on/reset default
value is 1800 seconds (3 minutes).
This command returns the current timeout value (in seconds) as an unsigned integer.
SYSTem:COMMunicate:SERial[1|2|3|4]:TRANsmit:AUTO
the transmit baud rate to be the same as that set for receive. When OFF, the transmit
baud rate may be set independently of the receive pacing mode. Setting the Transmit
baud rate with the
SYSTem . . . TRANSmit:BAUD <baud_rate>
automatically sets
SYSTem . . . TRANSmit:AUTO
ON.
SYSTem:COMMunicate:SERial[1|2|3|4]:TRANsmit:AUTO?
1) indicating the current state of transmit/receive baud rate linkage.
This command sets the baud rate (transmitter only) for the specified serial port. Valid
baud rates are: 75, 110, 150, 300, 600, 1200, 1800, 2000, 2400, 4800, 9600, 19200,
and 38400. All other values result in an error. Power-on/reset default baud rate is
9600.
This query returns the current transmitter baud rate for the specified port as an
unsigned integer.
SENSe:SERial[1|2|3|4]:TEXT?
command cannot be completed within the timeout
or
SOURce:SERial[1|2|3|4]:DATA?
to OFF. Power-on/reset default is
SCPI Command Reference
or
command
when set to 1 or ON, sets
command
returns an integer (0 or