set scan no<parameter>
set derive <parameter> [y | n]

5.4 UV Antifouling

Note that these commands apply only in the instance of a UV system installed on the
instrument.
Command
display uv
set uv time <ontime> <offtime>

5.5 Data Logging

Command
display status
set logmode [auto | manual]
logon
AML Oceanographic Ltd.
Note: Calculated parameters will output
-99.9999 if outside of the valid range of the
calculation formula.
Remove a parameter from being included in
the log file and monitor output.
Enable/disable a parameter to be calculated if
the required sensors are installed.
Derived Parameters:
depth (from pressure)
salc (salinity from conductivity)
salsv (salinity from sound velocity)
density (requires salinity)
sv (sound velocity, requires salinity)
Note: a parameter will only be output if it is
ALSO included in monitor scans. Refer 'SET
SCAN' command, above.
Description
Print the current UV settings.
Set the duty cycle of the UV lantern in seconds.
Note: Setting UV time to 0 0 will disable light emission.
Description
Print the current logging status, including in-water status,
number of samples recorded, log duration, and sample
rate.
By default logmode will be set to "auto" and will
automatically start logging when the water immersion is
detected.
Start logging when log mode is set to manual.