set burst monitor [samples | average]
display burst
Sleep Mode
The instrument can be configured to enter a low power state in between taking samples. The
instrument will only enter sleep mode if the sleep and sampling settings are configured so that
the sample period is longer than the minimum sleep time. By default the minimum time the
instrument is allowed to sleep is 40 seconds plus a 5 second delay after taking the sample.
Sleep mode is also compatible with Burst Mode.
Equation 2: Min Sleep Time + Sleep Delay
Equation 3: Min Sleep Time + Sleep Delay
Example 1:
Minimum Sleep Time = 40 s (default)
Sleep Delay = 5 s (default)
>SET SCAN SLEEP Y
>SET SAMPLE 2 MIN
45 < 120
Example 2:
>SET SCAN MODE BURST
>SET BURST INTERVAL 60
>SET BURST SAMPLES 30
AML Oceanographic Ltd.
Set whether the monitor output will display all
the burst samples or an average of the
samples.
Note: averaging applies only to monitor data.
Log files will record individual samples.
Display the current burst mode settings.
Burst Interval - Sample Duration
≤
-enable sleep mode
-set sample period greater than 45 seconds
-Equation 2 is satisfied and instrument will sleep between
taking samples
-enable burst mode
-set burst interval to 60 seconds
-set the number of burst samples to 30
Sample Period
≤
23/44