Monitor Modes - AML Oceanographic AML-6 RT User Manual

Multiparameter real time instrument
Table of Contents

Advertisement

>SET SAMPLE 5/S
45 < 60 - 30 / 5
= 45 < 54
Command
set scan sleep [y | n]
set minsleeptime <seconds>
set scan sleepdelay <seconds>

3.2.3 Monitor Modes

Behaviour of monitor functionality is configurable.
Command
set monitor robust [y | n]
set monitor auto <seconds>
set monitor startup [y | n]
AML Oceanographic Ltd.
-set sample rate fast enough so that burst interval - sample
duration is greater than minimum sleep time
-Equation 3 is satisfied and instrument will sleep between
bursts
Description
Enable/disable sleep mode.
Set the minimum number of seconds the
instrument is allowed to sleep. It is not
recommended to change this.
Default: 40 seconds
Min: 20 seconds
Set the number of seconds the instrument will
wait to sleep after taking a sample.
Default: 5 seconds
Description
Enable/disable robust monitor mode. When
enabled, monitoring is halted via 3 carriage
returns.
When disabled, a single carriage return will
halt monitoring.
Monitoring will begin automatically after an
interval of inactivity. Setting to "0" disables
auto monitor.
Min value: 5
This command applies only to monitoring via
serial connection
Enable/disable monitoring on startup. When