7 Diagnostics and servicing
26
6. Pull the cable out of the locking sleeve (4).
7. To avoid splicing of the steel cable, tin the cable before shortening
with a soldering iron and tighten the wire.
8. Shorten the cable with a cut-off wheel or a metal saw at the lower
end. Make sure the length is correct before shortening.
9. Shorten the insulation of the cable with a sharp knife by 60 mm
(2.36 in).
10. Shift the chamfered side of the locking sleeve (4) with a rotary
movement onto the cable.
Make sure that the cable is flush with the lower side of the locking
sleeve (4).
11. Lock the cable with the two pins (5).
Make sure that the cable shielding has no contact to the inner
steel cable.
12. Shift the isolating sleeve (6) onto the locking sleeve (4).
13. Shift the lower part of the weight (7) to the upper part of the
weight (1) and screw them together.
14. Lock the upper part of the weight (1) with pin (8).
15. Check the correct length of the probe.
16. Carry out a fresh adjustment. See chapter " Set up, new adjust-
ment".
7.5

How to proceed if a repair is necessary

You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
In case of repair, proceed as follows:
Print and fill out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.
VEGACAP 35 • Relay (DPDT)