4 .
DEVICE
USE
Install AtriCure
cryoICE
1.
Ensure ACM is properly connected to a N2O gas cylinder.
2.
The PROBE may be connected before the ACM has been turned on, while the ACM is being turned on, or
when the ACM is on and in READY Mode.
3.
Insert the corresponding connections on the pneumatic connectors as shown below in Figure 14. The sliding
ring will need to be manually pushed-in on the orange connector.
4.
Ensure each pneumatic connection is fully seated by listening for an audible "click" as each connector engag-
es its receptacle. Gently tug on each tube to ensure proper engagement with connector.
5.
Insert the corresponding red and black colored connections into the thermocouple connectors, see Figure 16.
6.
The PROBE icon, seen above in Figure 15, will extinguish if the PROBE is functioning properly and the approx-
imate room temperature will be displayed on the temperature display (typically 10 to 25° C [50°F to 77°F]).
An example of this is shown in Figure 17.
System Probe
Figure 14: Color Coded Pneumatic Connectors
Figure 15
Figure 17: Probe Temperature Display
Figure 16
9