Adjustment Procedures

DC CMRR

24
1. Set up the equipment as shown. (See Figure 12.)
a. Connect a BNC cable from the output of the 1103 to the Channel 1 input
of the oscilloscope. If the oscilloscope does not have 50 Ω input setting,
add a 50 Ω in-line BNC terminator at the oscilloscope input.
b. Connect the Y-lead adapter to the probe input.
c. Using square pins, connect both leads to the red binding post of a binding
post to BNC adapter.
d. Connect the BNC adapter to the function generator output. (If necessary,
use a BNC cable and a female-to-female BNC connector between the
BNC adapter and the generator.) If the function generator reference
(shield of the BNC connector) is isolated from ground, connect the ground
lead from the probe case to the black binding post.
NOTE.
Do not use the BNC to probe tip adapter for this connection. The power
level of the generator will exceed the terminator rating.
2. Connect a second BNC cable from the function generator Trigger or Sync
Output to the External Trigger or Channel 2 input of the oscilloscope.

Figure 12: Setup for DC CMRR and AC CMRR adjustments

3. Set the probe to ÷1 attenuation.
P6248 1.7 GHz Differential Probe Service Manual