TROUBLESHOOTING

Detector fails bump, fresh air or calibration:
1.
If fresh air fails, verify the filter is not clogged and the air at the inlet is not contaminated.
2.
Check if gas bottles are connected, sufficiently pressurized and contain the right gas.
(information on the bottle)
3.
Verify the station function with a known functional detector.
Detector fails alarm test:
1.
Redo the test
2.
Manually inspect the detector for LED, vibration and audio alarms
3.
Test a known fully functional detector in the station to verify station function
Station fail:
1.
Clear fault and retry
2.
Contact service representative