5.3 Removing transport locking screws
1
Transport locking screws
For the vacuum pump. Only present if the
instrument has a degasser or a CO
sor (MCS).
16
■■■■■■■■
Removing the transport locking screws
Figure 3
Removing the transport locking screws
Suppres-
2
Store the transport locking screws in a safe place. Reinsert the transport
locking screws each time you transport the instrument a significant dis-
tance.
CAUTION
The pumps may be damaged if you transport the instrument without
inserting the transport locking screws.
2
Transport locking screws
For the high-pressure pump.
930 Compact IC Flex Oven/Deg/MB (2.930.2180)
■■■■■■■■■■■■■■■■■■■■■■