Start-Up - KROHNE OPTIWAVE 1010 Supplementary Instructions Manual

Radar (fmcw) level transmitter for bypass chambers and magnetic level indicators (bm 26 advanced)
Hide thumbs Also See for OPTIWAVE 1010:
Table of Contents

Advertisement

OPTIWAVE 1010
WARNING!
Make sure that it is safe to supply electrical power. Do a start-up check:
• Are the wetted components (gasket, flange and antenna) resistant to corrosion by the tank
product?
• Does the data given on the device nameplate agree with the application and the process
conditons?
• Did you connect the equipotential bonding system correctly?
• Ex d applications:
Ex d applications: Are the cable entries, plugs and adaptors Ex d-approved?
Ex d applications:
Ex d applications:
• Ex t applications:
Ex t applications: Are the cable entries, plugs and adaptors Ex t-approved?
Ex t applications:
Ex t applications:
• Are you using an intrinsically-safe barrier within the correct parameters? For more data, refer
Ex ia equipment
to
maximum intrinsically-safe values.
• Did you install the correct cable entries? Is the terminal compartment correctly sealed?
• Make sure that the installation and wiring agree with local regulations.
WARNING!
Use the device only in the specified measuring range.
09/2015 - 4004191901 - AD ATEX OPTIWAVE 1010 R01 en
on page 13. The electrical circuit characteristics must not be more than the