EMS AIR-FLOW Operation Instructions Manual page 32

Prophylaxis system
Hide thumbs Also See for AIR-FLOW:
Table of Contents

Advertisement

Unclogging the handpiece using the EASY CLEAN
To get more force on the pressure use a small syringe (1 or 2 ml) filled with drink water.
Insert the assembled syringe with the EASY CLEAN into the handpiece lumina.
Press strongly if needed and maintain the pressure until the handpiece is unclogged (it can take 2 to 5
seconds) The water passing through the handpiece will dissolve completely the remaining powder inside the
lumina and ensure an optimal function.
After unclogging the handpiece, apply the periodical cleaning.
Unclog the handpiece before sterilizing.
26
FB-304_CN_rev_B_Syst_AF.indd 26
28/09/2018 14:43:40