2. The Flow phase will stop 60 seconds after voiding ends.
3. The Post Void Residual window (Figure 3-5) opens. Select the preferred method of capturing Post Void Residual
(PVR). PVR is the amount of fluid remaining in the bladder at the completion of the Voiding phase. Use of the
PVR functions is at the discretion of the supervising healthcare professional.
■ Drainage PVR allows the operator to pour drained fluid into the Urocap NXT to measure PVR.
■ Input PVR allows the operator to input the PVR measurement calculated by a separate measurement
device.
IMPORTANT
Do not empty the Urocap NXT beaker before adding PVR fluid drained from the patient's bladder. If the
beaker is filled to capacity, select the Drainage PVR or Input PVR options.
4. In the Post Void Residual dialog box, click OK to save data and advance to the report. Clicking Cancel abandons
any information you input and closes the Post Void Residual dialog box.
NOTE
Post Void Residual can also be entered and edited in the Report.
26
Figure 3-4: FreeFlow Record Phase
Figure 3-5: The Post Void Residual Dialog Box
Laborie NXT Go Owner's Manual