Additional information

IPROXX 2
The following standards and norms were used in the versions with at least IPX4 and higher
in according to
-
EN 60601-1:2006 + A1:2013, IEC 60601-1:2005 + A1:2012 (short description: Edition 3.1), Med-
ical electrical equipment.
-
EN 60601-1-2:2015, IEC 60601-1-2:2014 (short description: Edition 4.0), EMC
IEC/EN 60601-1, section 7.8.1
IEC/EN 60601-1, section 8.10
IEC/EN 60601-1, Abschn. 8.10.4.1
IEC/EN 60601-1, Abschn. 9.3
IEC/EN 60601-1, Abschn. 11
IEC/EN 60601-1, Abschn. 13
IEC/EN 60601-1, Abschn. 15.4.7.1
IEC/EN 60601-1, Abschn. 15.4.7.2
IEC/EN 60601-1, Abschn. 16.6
IEC/EN 60601-1, Abschn. 17
The following standards and norms were used in accordance with
-
EN 60601-2-52, IEC 60601-2-52, (Particular requirements for the safety and essential perfor-
mance of medical beds), the following standards have been used
IEC/EN 60601-2-52, Section 201.3.8
IEC/EN 60601-2-52, Section 201.9.2.2.5
IEC/EN 60601-2-52, Section 201.9.2.3.1
IEC/EN 60601-2-52, Section
201.11.6.5.101
IEC/EN 60601-2-52, Section 201.11.1.1
IEC/EN 60601-2-52, Section
201.15.3.4.1
IEC/EN 60601-2-52, Section 201.15.4.4
IEC/EN 60601-2-52, Section BB.3.3.3
IEC/EN 60601-2-52, Section BB.3.4.1
Colours of signal lights
Components and wiring
Operating voltage / Input voltage
Hazards due to surfaces, corners and edges
Temperatures
Hazardous situation and fault conditions: 15W limited
supply circuit
Mechanical strength
Accidental actuation
Leakage currents: touch current <100μA
EMC
Applied Part B
Hold to run controls / momentary contact switch
Unintentional movement:
Means to deactivate IPROXX 2 SM, IPROXX 2 SM+,
IPROXX 2 SMP
Ingress of water: min IPX4
Temperature
Falling test
Indicator lights "Ready for Normal Use" not required
Dimensions for hand grips
Operating forces