Changing the Voltage Setting
1. Open fuse holder compartment with a screwdriver (Figure 12).
2. With the screwdriver lever out the fuse holder (Figure 13).
3. Replace the two fuses according to the table below:
4. Insert the fuse holder keeping the set voltage label at the top.
5. Close the fuse compartment by pressing gently.
6. If the fuse holder is inserted correctly you can read the voltage
selected.
4titude Ltd, The North Barn, Surrey Hills Business Park, Damphurst Lane, Wotton, Dorking, Surrey, RH5 6QT, UK
Tel: +44 (0) 1306 884885 Fax: +44 (0) 1306 884886 info@4ti.co.uk www.4ti.co.uk
Power Voltage
115V
230V
Table 1 - Fuse ratings for 4s2
(EN 60127)
3.15A (T)
2A (T)
Page | 17