Digit flexdmm calibration procedure
Hide thumbs Also See for PXI-4070:
Table of Contents

Advertisement

Frequency

Calibration Function Reference

Technical Support Resources

NI Web Support

© National Instruments Corporation
NI 6608
Output Frequency
1 Hz
20 kHz
500 kHz
For detailed information about the NI-DMM calibration functions used in
this procedure, refer to the VI Reference or the C/C++/VB Function
Reference topics of the NI Digital Multimeters Help, located at Start»
Programs»National Instruments»NI-DMM»Documentation.
National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer's
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
supports calibration, you can obtain the calibration certificate for your
product at
ni.com/calibration
Table 17. Frequency Limits
2-Year Limits
Lower
0.9999 Hz
19.998 kHz
499.95 kHz
ni.com
ni.com/hardref.nsf
.
63
Upper
1.0001 Hz
20.002 kHz
500.05 kHz
.
. If your product
NI PXI-4070 Calibration Procedure