Adjusting Current Modes - National Instruments PXI-4070 Manual

Digit flexdmm calibration procedure
Hide thumbs Also See for PXI-4070:
Table of Contents

Advertisement

Calibrator Output
Iteration
Amplitude
39
50 V
50 V
40
10 V
10 V

Adjusting Current Modes

NI PXI-4070 Calibration Procedure
Table 10. niDMM_CalAdjustACFilter Parameters (Continued)
Frequency
(kHz)
300
NIDMM_VAL_AC_VOLTS
300
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
500
NIDMM_VAL_AC_VOLTS
500
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
9.
Reset the calibrator for safety reasons.
10. Call
niDMM_CalAdjustMisc
NIDMM_EXTCAL_MISCCAL_SECTION
You have completed adjusting the AC voltage modes of the NI 4070. Select
one of the following options:
If you want to perform additional adjustments, go to the Adjusting
Current Modes section.
If you do not want to perform the current mode adjustments, go to the
Verification Procedures
coefficients before saving them to the EEPROM.
If you do not want to verify the adjustments you have just made, go to
the
Completing the Adjustment Procedures
If you do not use the current modes (DC and AC), or the accuracy is
insignificant for your application, you can skip this section and select one
of the following options:
If you skip this section and you want to verify the new calibration
coefficients before saving them to the EEPROM, repeat the
Verification Procedures
Equipment).
If you skip this section and you do not want to verify the new
calibration coefficients, go to the
Procedures
section.
niDMM_CalAdjustACFilter Parameters
Mode
with Type set to
.
section to verify your new calibration
section (except for
Completing the Adjustment
54
Range
Frequency
(V)
(Hz)
300
300e3
300
300e3
300
500e3
300
500e3
section.
Setting Up the Test
ni.com