Verifying AC Voltage

Caution
Step 2 must be performed correctly to avoid shorting or electrical shock from
high voltage on unconnected spade lugs.
© National Instruments Corporation
You have completed verifying the DC voltage of the NI 4070. Select one of
the following options:
If you want to continue verifying other modes, go to the Verifying AC
Voltage section.
If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the
Procedures
section.
If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
the session.
To verify AC voltage of the NI 4070, complete the following steps:
1.
Reset the calibrator.
2.
Connect the calibration cable assembly with the connector block
plugged into the NI 4070 and with the spade lugs connected to the
appropriate binding posts on the calibrator, as described in Figure 3
and Table 1.
3.
Output 5 mV at 1 kHz on the calibrator.
4.
Call
niDMM_reset
5.
Call
niDMM_ConfigureMeasurement
parameters:
Function =
NIDMM_VAL_AC_VOLTS
Range =
0.05
Resolution =
6.
Call
niDMM_Read
limits listed in Table 12.
7.
Call
niDMM_ConfigureMeasurement
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range =
0.05
Resolution =
8.
Call
niDMM_Read
limits listed in Table 12.
9.
Output 50 mV at 30 Hz on the calibrator.
Completing the Adjustment
to reset the NI 4070 to a known state.
50e–9
. Verify that this measurement falls between the
50e–9
. Verify that this measurement falls between the
19
to close
niDMM_close
with the following
with the following
NI PXI-4070 Calibration Procedure