Dc Current; Ac Current - National Instruments PXI-4070 Manual

Digit flexdmm calibration procedure
Hide thumbs Also See for PXI-4070:
Table of Contents

Advertisement

DC Current

AC Current

NI PXI-4070 Calibration Procedure
Table 15. NI 4070 DC Current Verification Limits
Calibrator
Amplitude
Range
20 mA
20 mA
–20 mA
20 mA
0 A
20 mA
0 A
200 mA
0 A
200 mA
200 mA
–200 mA
200 mA
1 A
–1 A
Table 16. NI 4070 AC Current Verification Limits
Calibrator Output
Amplitude
Frequency
1 mA
1 kHz
10 mA
1 kHz
10 mA
1 kHz
100 mA
1 kHz
100 mA
1 kHz
1 A
1 kHz
62
Lower
19.9905 mA
–20.0095 mA
–1.5 µA
–4 µA
–20 µA
1 A
199.916 mA
–200.084 mA
1 A
0.99948 A
1 A
–1.00052 A
Range
10 mA
0.9976 mA
10 mA
9.994 mA
100 mA
9.976 mA
100 mA
99.94 mA
1 A
1 A
2-Year Limits
Upper
20.0095 mA
–19.9905 mA
1.5 µA
4 µA
20 µA
200.084 mA
–199.916 mA
1.00052 A
–0.99948 A
2-Year Limits
Lower
Upper
1.0024 mA
10.006 mA
10.024 mA
100.06 mA
99.7 mA
100.3 mA
0.9988 A
1.0012 A
ni.com