6.4 Chiller (May 2020)

● The Q-SUN Xe-3 will determine if the chiller is needed during each step of a test program, based on the type of step,
the temperature set point, and the lab temperature.
● Make sure the chiller power cable is connected to the power outlet on the rear of the Xe-3 (see
● IMPORTANT: The ON/OFF switch, located inside the chiller front door
● The chiller may not start running immediately when the switch is turned to ON, as the Q-SUN will turn the chiller ON
and OFF as needed for the step being run.
● The damper position must be set to the lowest temperature expected in the room where the Q-SUN is operating
(Figure
6.4a).
● Using the graduations on the damper as a guide, the damper louvers should be adjusted
if the room is 20
C (68
o
● In a room of 10
C (50
o
OFF Switch
Damper
Adjustment
F) or warmer.
o
F) the damper should be almost closed.
o
Chiller ON/
Figure 6.4a: Chiller ON/OFF switch and damper location.
Figure 6.4b: Damper adjustment location at top rear of chiller.
(Figure
6.4a), should remain in the ON position.
(Figure
Damper
Section 6. Start Up
Figure
5.8h).
6.4b) to be fully open
51
51