How to Notify Service if the Recorder is Defective

If the recorder is defective, it will be detected during the self-test when the
recorder is switched on. In case a defect is detected during the self-test,
the following sequence of LED flashes will alert you that a problem has
been detected:
Both LED flash yellow.
Both LED flash green.
A two-second pause follows.
The green Event Button LED flashes.
If any of these failures occur, contact GE Technical Support. By counting
the number of flashes, service personnel can determine the cause of the
defect.
Refer to the following table for the defect identified by the number of LED
flashes.
2067634-077 Revision B
Enter the Administrator Password in the second line.
Click Login.
The Hookup tab opens.
Change to the Devices tab.
Select the recorder.
If the recorder to be reset is not listed, click the M symbol.
Once the recorder to be reset is listed, select it with a single click
and click Reset. A message confirms that the recorder has been
reset.