Bluetooth Troubleshooting - GE SEER 1000 Operating Manual

Ecg recorder and mobile application
Table of Contents

Advertisement

Message
"Wrong password."
"Wrong technician or password."

Bluetooth Troubleshooting

Trouble Indicators
Error while connecting
recorder and app
After starting a record-
ing, no wireless connec-
tivity active
2067634-077 Revision B
Cause
Recorder and app device are
either too close to or too far
away from each other.
The app device's Bluetooth
connectivity is not enabled.
Objects between recorder
and application device affect
the connection.
RF interfering devices affect
the connection.
The recorder's Bluetooth
connectivity is not enabled
because this is switched off
after the start of a recording.
SEER™ 1000
Recommendation
The password is wrong. Try a valid pass-
word.
The technician ID does not exist or the
password is wrong. Try a valid combina-
tion of technician ID and password.
Recommendation
For best results, do not position
the devices closer than 50 cm
(approx. 20 inches) to each
other or more than 10 m (ap-
prox. 11 yards) from each
other.
Verify that your device's Blue-
tooth connectivity is enabled.
Remove any objects from the
line of sight between recorder
and application device.
Remove any possible RF interfer-
ing devices from the vicinity of re-
corder and application device.
Press and hold the Event But-
ton for approx. 3 seconds until
the multi-beep indicates wire-
less connectivity is activated.
Troubleshooting