Electrode Placement And Waveform Preview - GE SEER 1000 Operating Manual

Ecg recorder and mobile application
Table of Contents

Advertisement

Using the Apple iOS App
4.

Electrode Placement and Waveform Preview

WARNING
WAVEFORM DIAGNOSTIC QUALITY – The ECG waveforms shown on
the preview screen are used for assessing the quality of the connec-
tions only.
Do not use these preview waveforms for diagnostic purposes.
CAUTION
SIGNAL QUALITY – If there is a severe muscle interference or artifact
on the ECG when you tap the top of the electrodes lightly or whe n the
patient moves, it is possible that the electrodes are not attached properly.
To increase the accuracy of analysis, make sure that the channel has
a high amplitude of QRS complex. If the amplitude is small, shift the
electrodes to find a suitable location for electrode attachment.
The app will display a torso image with the recommended electrode placement.
The system will automatically detect the cable type and display the appropriate
configuration. Double-tap the torso to zoom in or to zoom out again.
64
The recorder identifier will now appear in the device list.
Select the recorder to move to the overview screen. Refer to the
"Electrode Placement and Waveform Preview" section on page
64 for next steps.
NOTE
If the recorder is in the process of recording or contains a re-
cording, a warning is displayed.
SEER™ 1000
2067634-077 Revision B