QUICK REFERENCE GUIDE
IT IS ESSENTIAL THAT YOU READ AND UNDERSTAND THE
OPERATOR'S MANUAL BEFORE USING THIS DEVICE
Registered in Accordance with ISO 13485

Advertisement

loading
Need help?

Need help?

Do you have a question about the RAPID INFUSER RI-2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Belmont RAPID INFUSER RI-2

  • Page 1 QUICK REFERENCE GUIDE IT IS ESSENTIAL THAT YOU READ AND UNDERSTAND THE OPERATOR’S MANUAL BEFORE USING THIS DEVICE Registered in Accordance with ISO 13485...
  • Page 2: Pole Mounting

    Pole Clamp Release Handle approximately 30” from the ground. • Mount The Belmont ® Rapid Infuser RI-2 RI-2 UNIT above the Pole Mount Support Assembly. • Push the “Pole Clamp Release Handle” down to lock the RI-2 onto the IV Pole.
  • Page 3 Plug the main connector of the detachable power cable into a dedicated circuit breaker. Fully seat the device connector Quick Setup ® (C-19) of the power cable into the power receptacle on the back of The Belmont Rapid Infuser RI-2. If a moisture guard is Guide present, ensure it is over the device connector and flat against the back of the machine.
  • Page 4 REPLACING RESERVIOR CHAMBER / INSTALLING 3 LITER RESERVOIR Replace Filter / Standard Reservoir Chamber (Fluid Supply) Install 3 Liter Reservoir Reservoir Chamber Clamp this off before removing the reservoir chamber Disconnect luer lock and quick connector to Connect 3.0 Liter Reservoir to remove the chamber the 3-Spike Disposable Set REPLACING RESERVOIR CHAMBER...
  • Page 5: Operational Alarms

    OPERATIONAL ALARMS ALARM / MESSAGE POSSIBLE CAUSES SOLUTION UNIT DOES NOT POWER UP • Power switch not completely depressed • Depress power switch completely AIR DETECTION • Air in the line • Open door, squeeze tubing directly below air detector sensor to clear air. •...
  • Page 6 Call technical support SYSTEM PARAMETERS SETTING The CALIBRATION/SET-UP screen is accessed by pressing SERVICE key during power-up This key appears on the BELMONT logo screen and remains active for 4.5 seconds before system enters the PRIME screen. DATE/TIME DISPLAY BRIGHTNESS •...
  • Page 7 Rapid Infuser Incompatible Solutions Solution Description Compatible? Conversions & Heating Sodium Bicarbonate Solutions Heats to 39° C Heats to 37.5° C mL/min mL/hr mL/min mL/hr ½ NS 0.45% NaCl 3600 3% NS 3% NaCl 4200 Platelets Should not be diluted, stick to tubing 4800 Cryoprecipitate Should not be diluted...