Ecleris MINIVAC User Manual

Ecleris MINIVAC User Manual

Microdermabrasion and vacuum therapy

Advertisement

Quick Links

Version: 2019.11.01
All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be
reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it
be reproduced, multiplied or published through any electronic data processing form.
The information contained in this manual may be modified or broaden without previous notification, without it representing any
responsibility to ECLERIS.
ECLERIS S.R.L.
FRANCISCO N. LAPRIDA 4955. VILLA
MARTELLI
(B1603ABK)
ARGENTINA
Tel/Fax: +54 11 48383200
Web: www.ecleris.com
E-Mail:
Directora Técnica: Farm. Doris Inés Rasi
Sanchez, MN: 12.856
Autorizado por la ANMAT PM-1120-43
Condición de venta: "Venta exclusiva a
profesionales e instituciones sanitarias".
USER MANUAL
MICRODERMABRASION AND VACUUM THERAPY
ABOUT THIS MANUAL
Year: 2019
BUENOS
AIRES.
info@ecleris.com
MINIVAC
Language: English
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº18, CP29006,
Málaga, Spain
Ph. +34951214054
Email: info@cmcmedicaldevices.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MINIVAC and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Ecleris MINIVAC

  • Page 1 All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it be reproduced, multiplied or published through any electronic data processing form.
  • Page 2: Table Of Contents

    USER MANUAL – MINIVAC TABLE OF CONTENT SYMBOLS ..........................3 SAFETY INFORMATION ......................3 WARNINGS .......................... 3 CONTRAINDICATIONS ....................... 4 DESCRIPTION......................... 5 INTRODUCTION AND FUNCTIONING ................5 INTENDED USE ........................5 INTENDED USER ........................ 6 COMPONENTS ........................6 3.4.1. MAIN UNIT ..........................6 INSTALLATION AND START UP .....................
  • Page 3: Symbols

    Should you have any doubt regarding the contents of this manual, the operation or equipment safety, please contact your local distributor or the nearest ECLERIS office. ECLERIS shall not be liable for the damages that may occur if the equipment is not used according to these instructions and following the criteria mentioned herein below: ...
  • Page 4: Contraindications

     Do not open the device. The diagnose, service and repairs must be performed by ECLERIS, or a company authorized by ECLERIS. ECLERIS will not be liable for the damage caused by repairs, or service performed by unauthorized personnel. ...
  • Page 5: Description

    3. DESCRIPTION 3.1 INTRODUCTION AND FUNCTIONING The Ecleris Minivac has been designed for cosmetic applications. It constitutes an aspiration system with vacuum control for skin exfoliation. Microdermabrasion is a technique that removes the superficial layers of the skin through a mechanic and controlled exfoliation performed with the Diamond Tip kit included.
  • Page 6: Intended User

    3.3 INTENDED USER Is expected to be used by health professionals dermatologists or experienced beautician. 3.4 COMPONENTS The Ecleris Minivac includes the following necessary components. Please identify and check them before you proceed with the installation:  1 Main unit ...
  • Page 7: Installation And Start Up

    Green Purple 8) Connect the AC interlock cable to the Ecleris Minivac. Check that the electrical network voltage to be used is the same as the one stated in the voltage selector switch on rear panel of the device and that the correct fuse is installed.
  • Page 8: Vacuum System Verification

     Check that all piping is in good conditions and not strangulated. If before starting the Ecleris Minivac you detect any abnormality, please contact the technical department of your local distributor prior to using it. Connect the AC interlock cable to power supply. Press the master ON/OFF switch to turn on the Ecleris Minivac.
  • Page 9: Diamond Tip

    USER MANUAL – MINIVAC 3) Suction will start as soon as the Diamond Tip gets in touch with the skin and will stop immediately when your finger is lifted from the hole on the top of the vacuum regulator. This allows the operator to exert proper treatment control.
  • Page 10: Post-Treatment Recommendations

    Make-up can be used immediately, but preferably not for the first 24 hours. 6. MAINTENANCE 6.1 CLEANING Before cleaning the Ecleris Minivac, disconnect the power supply cable from the electrical network. 6.1.1. EXTERNAL SURFACES Clean the external surface of the equipment using a piece of cotton or cloth moistened with mild detergent or soap.
  • Page 11: Micro-Porous Filter

    6.2 FUSE CHANGE To avoid time losses due to downtime of the Ecleris Minivac, it is recommended to have a spare fuse. 1) Disconnect the power supply cable from the electrical network and the equipment.
  • Page 12: General Considerations

    Faulty connection between the cable and the plug.  Oxide corrosion in connection pins 6.4 GENERAL CONSIDERATIONS To perform any revision or maintenance work on the Minivac, contact your local distributor or the nearest Ecleris office. 7. TECHNICAL SPECIFICATIONS Minivac models...
  • Page 13 Place the device on a vibration- vibration-free surface. free surface. The compressor is not Compressor overheating due to Turn the Ecleris Minivac off for 5 working over voltage minutes. Check that the selected voltage is the correct one and turn it on again...
  • Page 14: Technical Service

    USER MANUAL – MINIVAC If the problem cannot be solved, please contact your local distributor or the closest ECLERIS office and request technical service. 9. TECHNICAL SERVICE The Ecleris Minivac should be repaired and/or maintained exclusively by ECLERIS trained and qualified personnel in authorized installations and the only parts repaired should be the ones that ECLERIS states as repairable.