Shodex OHpak SB-2000 Series Operation Manual

Advertisement

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.
1. Introduction
Thank you for purchasing the Shodex product. Shodex OHpak SB-2000 series is size exclusion
chromatography columns used with aqueous solvents. The column series is suitable for preparative isolation of
various water-soluble polymers, proteins, and oligomers. Please select columns with target molecular weight
ranges (see below) that meet the molecular weight range of your samples.
2. Column Components
Please refer to the Shodex website: https://www.shodex.com/en/da/07.html
3. Column Specifications
Product
Product Name
Code
OHpak SB-2002
F6516011
F6516012
OHpak SB-2002.5
F6516013
OHpak SB-2003
F6516014
OHpak SB-2004
F6516015
OHpak SB-2005
OHpak SB-2006
F6516016
OHpak SB-2006M
F6516017
OHpak SB-G 8B
F6709555
SB-2006M is a mixed-gel column, provides a wider molecular-weight analysis range.
Operation Manual
Shodex OHpak SB-2000 series
Column Size (mm)
I.D
Length
20.0
300
20.0
300
20.0
300
20.0
300
20.0
300
20.0
300
20.0
300
8.0
50
Resonac Corporation (https://www.shodex.com/)
Theoretical
Particle Size
Plate Number
(μm)
(Per Column)
15
≥ 9,000
10
≥ 12,000
10
≥ 12,000
18
≥ 12,000
20
≥ 12,000
20
≥ 12,000
20
≥ 12,000
18
(Guard Column)
Reference value only / Measured with pullulan
*1
ver.20221202E
Exclusion Limit
*1
1,000
10,000
100,000
1,000,000
(4,000,000)
*2
(20,000,000)
*2
(20,000,000)
*2
-
Estimated value
*2
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OHpak SB-2000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex OHpak SB-2000 Series

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex OHpak SB-2000 series is size exclusion chromatography columns used with aqueous solvents. The column series is suitable for preparative isolation of various water-soluble polymers, proteins, and oligomers.
  • Page 2 · Do not use borate buffer as it forms a complex with diol group of the packing material. · Column pressure is influenced by eluent composition, flow rate, and column temperature. When changing the eluent compositions, adjust the flow rate and column temperature so that the column pressure remains below the usable maximum pressure.. Resonac Corporation (https://www.shodex.com/)
  • Page 3: Eluent Preparation

    · When a sample is dissolved in a solvent other than the eluent and if the sample matrix contains components which do not dissolve in the eluent fully, precipitates may form and clog the column. Note · Use of a guard column is recommended to protect the analytical column. Resonac Corporation (https://www.shodex.com/)
  • Page 4: Column Installation

    Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. Resonac Corporation (https://www.shodex.com/)
  • Page 5: Column Storage

    Securely tighten the end caps and store the column at a location with stable temperature (a cool and dark space is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. Attention! · Never allow inside the column to dry. It can damage the column. Resonac Corporation (https://www.shodex.com/)
  • Page 6 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...