Shodex Sugar SZ5532 Operation Manual

Shodex Sugar SZ5532 Operation Manual

Advertisement

Quick Links

(Please read this manual carefully to achieve the best and consistent column performance for a long time)

Important Handling Instructions

Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column name tag, and
enclosed Certificate of Analysis (CoA) are matching and correct.

1. Introduction

Thank you for purchasing the Shodex product. Shodex SUGAR SZ5532 is a high performance
chromatography column which is specifically designed for the separations and analysis of sugars. A strong acid
cation exchange type packing gel, prepared from a rigid resin of styrene divinylbenzene copolymer provides a
mixed separation mode. The analytes are separated by a combination of ligand exchange and hydrophilic
chromatography (HILIC) modes. It is capable of separating and analyzing disaccharides or trisaccharides from
each other.

2. Column Components

3. Column Specifications

Product
Product Name
Code
F7001300
SUGAR SZ5532
F6700110
SUGAR SZ-G
Base Material
: Spherical porous particles of styrene divinylbenzene copolymer modified with sulfo group
(counter ion: Zn
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water/Acetonitrile = 30/70
Operation Manual
Shodex
SUGAR
Column Size (mm)
I.D.
6.0
4.6
2+
)
SHOWA DENKO K.K. (https://www.shodex.com/)
SZ5532
Particle
Size (μm)
Length
150
6
10
6
ver.210629E
Theoretical Plate Number
(Per Column)
≥ 5.500
(Guard Column)
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sugar SZ5532 and is the answer not in the manual?

Questions and answers

Summary of Contents for Shodex Sugar SZ5532

  • Page 1 Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex SUGAR SZ5532 is a high performance chromatography column which is specifically designed for the separations and analysis of sugars. A strong acid cation exchange type packing gel, prepared from a rigid resin of styrene divinylbenzene copolymer provides a mixed separation mode.
  • Page 2 (7) When a sample contains a large amount of organic acids, use an OH-form anion exchange resin to remove them. (8) When a sample contains hydrophobic substances or surfactants, use a reversed-phase solid phase extraction to remove them. Note · Use of guard column is recommended to protect the analytical column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 3 When replacing the solvent, set the column temperature at 50 °C and start the system at 0.2 - 0.3 mL/min. Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume. Attention! · Frequent eluent replacement may damage the column, and thus not recommended. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 4 (a cool and dark space is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. Attention! · Never allow inside of the column to dry. It can damage the column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 5 (2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
  • Page 6 11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.

Table of Contents