This section explains the operating environment for the ZR-RX45.

Ambient Operating Conditions

(1)
Ambient temperature and humidity (the ZR-RX45 must be operated within the following ranges.)
• Temperature range: 0 to 45°C (0 to 40°C when a battery pack is mounted.)
• Humidity range: 5 to 85% RH
(2)
Environment (do not use in the following locations.)
• A Location such as being exposed to direct sunlight
• Locations exposed to salty air, corrosive gases, or organic solvents
• Dusty locations
• Locations subject to vibration or impact
• Locations subject to voltage surge or electromagnetic interference such as lightning or electric furnaces
(3)
Installation category (over-voltage category)
• The ZR-RX45 belongs to Installation Category II defined in IEC60664-1.
• Never use the ZR-RX45 for Installation Category III or IV.
(4)
Measurement category
• The ZR-RX45 belongs to Measurement Category I defined in IEC61010.
• The ZR-RX45 cannot be used for Measurement Category II, III, or IV.
N o te
If condensation occurs...
Condensation occurs in the form of water droplets on the device surfaces and interior when the ZR-RX45 is moved
from a cold to a warm location. Using the ZR-RX45 with condensation will cause malfunctioning.
Wait until the condensation has disappeared before turning on the power.

Warming-up Before Use

The ZR-RX45 should be allowed to warm up with the power turned on for approximately 30 minutes to ensure
that it operates according to the specified performance.

Configuration When in Use

Do not use the ZR-RX45 standing upright or at an angle. It must always be laid flat.
Usage Configuration
18

Operating Environment

Laid flat