Page 2
ZR-RX45. When using the ZR-RX45, be sure to observe the following: · The ZR-RX45 must be operated by personnel knowledgeable in electrical engineering. · To ensure correct use, please read this manual thoroughly to deepen your understanding of the product.
APPLICATION CONSIDERATIONS (Please Read) General Description User's Manual Checks and Preparation Settings and Measurement SPECIFICATIONS APPENDIX Portable Multi Logger ZR-RX45...
Page 4
LOSS OF PROFITS OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR STRICT LIABILITY. In no event shall responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.
Page 5
Performance data given in this document is provided as a guide for the user in determining suitability and does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users must correlate it to actual application requirements. Actual performance is subject to the OMRON Warranty and Limitations of Liability.
Page 6
Indicates the possibility of electric shock under specific conditions. Indicates prohibition when there is a risk of minor injury from electrical shock or other source if the product is disassembled. Indicates general prohibitions for which there is no specific symbol. ZR-RX45 User’s Manual...
Page 7
Fire or hazard may occur in rare occasions from ignition, rupture or combustion. Do not use battery packs other than ZR-XRB1. Hazard may occur from electric shock. Be sure to connect the terminal of this product to the cable first, and connect the measurement object. ZR-RX45 User’s Manual...
Page 8
• The ZR-RX45 conforms to the IEC60664-1 installation category II, and must not be used under the environment of the installation category III and IV. 4.Measurement category • The ZR-RX45 is classified as measurement category I defined by IEC61010-1, and must not be used within measurement category II, III and IV. 5.Others •...
Page 9
• Do not drop or apply strong impact or force to the product. It may cause malfunction of the monitor or the main unit. 6.Maintenance • Do not use thinner, benzine, acetone or kerosene to clean this product. • Calibration should be performed periodically to maintain measurement accuracy. ZR-RX45 User’s Manual...
Page 10
• Specal PC software “Wave Inspire RX“ (tryout) • Basic PC software “Smart Viewer RXW” • User's Manual PDF files (this manual) • “Wave Inspire RX“ Software Manual PDF files • “Smart Viewer RXW” Software Manual PDF files ZR-RX45 User’s Manual...
Page 11
Editor's Note Meaning of Symbols Menu items that are displayed on the ZR-RX45's LCD screen, and windows, dialog boxes and other GUI elements displayed on the PC are indicated enclosed by brackets "[ ]". Visual Aids Indicates points that are important to achieve the full product performance, Important such as operational precautions.
Page 17
General Description This chapter provides a general description of the ZR-RX45 and its features. Overview Features Operating Environment Notes on Temperature Measurement Notes on Using the Monitor Changing the Display Language...
ZR-RX45 are provided with 20 channels as a standard measurement feature, or can be extended up to 200 channels by attaching additional terminal sets. ZR-RX45 are also equipped with an internal flash memory to store data and enable the direct capture of a large volume of data to USB memory.
• The USB drive mode function enables the ZR-RX45's internal memory to be recognized as an external drive by your PC. (Connect the ZR-RX45 to your PC and turn on the power supply to the ZR-RX45 while holding down the [START] key.)
N o te If condensation occurs... Condensation occurs in the form of water droplets on the device surfaces and interior when the ZR-RX45 is moved from a cold to a warm location. Using the ZR-RX45 with condensation will cause malfunctioning.
Page 21
Important Do not block the air vent on the ZR-RX45, as this will cause malfunctioning. Measurement accuracy may not be satisfactory if the system is used in a condition other than described above. ZR-RX45 User’s Manual Operating Environment...
Please observe the following precautions when performing temperature measurement. • Do not block the air vents. Always provide a space of at least 30 cm on all sides of the ZR-RX45. • For stabilized temperature measurement, allow the ZR-RX45 to warm up for at least 30 minutes after turning it on.
If the screen saver operates, press any key to restore the display. Important • Condensation may form on the LCD screen if the ZR-RX45 is moved from a cold to a warm location. If this occurs, wait until the LCD screen warms up to room temperature.
Changing the Display Language You can choose the language displayed on the screen. The default display language is set to English when the ZR-RX45 is shipped overseas. To change the display language, see the instructions in "OTHER:Language". Changing the Display Language...
Checks and Preparation This chapter explains how to check the ZR-RX45's external casing andaccessories, and how to prepare the ZR-RX45 for operation. Part Names and Functions Connecting the Power Cable and Turning on the Power Connecting the Signal Input Cables...
Part Names and Functions This section describes the names and function of parts of the ZR-RX45. Operation status LED PC interface terminals Monitor Power switch • USB • POWER : ON when the power is ON • START : ON during data capture •...
Plug the AC cable into the AC adapter. AC adapter AC cable Connect the output side of the AC adapter to the connector on the ZR-RX45. AC adapter cable ZR-RX45 User’s Manual Connecting the Power Cable and Turning on the Power...
Page 28
Press the power switch on the ZR-RX45 to the ON side to turn on the power. Important Always connect the GND terminal and refer to the safety precautions. The ZR-RX45 must be grounded even when connected to other devices and sharing a common ground level.
Important Be sure to check the polarity of the wire tips when performing wiring. Press the power switch on the ZR-RX45 to the ON side to turn on the power. ZR-RX45 User’s Manual Connecting the Power Cable and Turning on the Power...
......... Dedicated terminal when connecting resistance temperature detector * Resistance temperature detector input terminals A (+) and B (-) are isolated within each channel. Terminal b is shorted within all channels. Connecting the Signal Input Cables ZR-RX45 User’s Manual...
Page 31
Filter operation is on a moving average basis. The average value of the set sampling count is used. If the sample interval exceeds 30 seconds, the average value of data obtained in a sub-sample (30 seconds) is used. ZR-RX45 User’s Manual Connecting the Signal Input Cables...
Approx. 0.5 V (+2.5 to +3 V) Alarm Output Specifications Item Description Number of Output channels Output format Open collector output +5 V, 10 K pull-up resistance See the next page for details on alarm output. Logic Alarm Cable Connection and Functions ZR-RX45 User’s Manual...
Page 34
White with black dotted line Alarm output Yellow with red dotted line Yellow with black dotted line Pink with red dotted line Trigger input/sampling input Pink with black dotted line Shielded Logic Alarm Cable Connection and Functions ZR-RX45 User’s Manual...
Attach the USB memory to the USB memory terminal. USB Memory Important When you attach the USB memory to ZR-RX45, be careful during handling so as not to bump or drop the unit. <Specifications of supported USB memory> • Power source : +5 V •...
Connecting to a PC Use the USB, LAN Interface to connect the ZR-RX45 to a PC Connection Using a USB Cable Use the USB cable to connect the ZR-RX45 to a PC. USB cable N o te If the USB cable is used, the USB driver must be installed in your PC. Please refer to "Installing the USB Driver" in the "PC Software Manual".
LAN Connection Use a LAN cable to connect the ZR-RX45 to a PC. LAN cable Cable Types • Use a crossing cable when connecting directly to a PC, without using a hub. LAN cable (crossing) • Use a straight cable when using a hub.
Using the Battery Pack (ZR-XRB1: Option) • The ZR-XRB1 (option) is the only battery type that can be used with the ZR-RX45. • Refer to the specifications (P.133) for information on the battery run time. • The operating temperature range of the ZR-RX45 with a battery pack mounted is as follows: Running on battery : 0 to 40°C...
CHARGE LED N o te • ZR-RX45 is equipped with a temperature monitor function which starts automatic charging as soon as it is cooled down. Therefore, depending on the internal temperature, charging may not be performed immediately. • When charging is attempted while the power is ON, charging may not be performed immediately even if the tem- perature environment conforms to the specification.
Connecting the Humidity Sensor (Option) Connect the + and - lead wires of the humidity sensor (the ZR-XRH1 option) to the desired terminals, and then insert the round connector into the 5V OUT connector on the ZR-RX45. Connected to the 5V OUT terminal.
Remove and mount terminal units as shown below. Important Make sure the ZR-RX45's power is OFF when removing or mounting terminal units. To Remove Pull the terminal unit out towards the direction indicated by the arrow while pressing the two locks at the bottom of the unit.
To Mount Insert the tabs at the top of the terminal unit into the slots of the ZR-RX45, and push in the unit until the lock tabs at the bottom of the unit are securely locked. (1) Insert tabs at the top of the terminal unit into the slots.
Mounting the Extension Terminal Base Set (ZR-XRE1) Mount the extension terminal base set as shown below. Important Make sure the ZR-RX45's power is OFF when mounting the extension terminals. Set Contents Extension Terminal Base Unit : 1 Extension Terminal Cable : 1 To Mount Remove the terminal unit mounted to the ZR-RX45 (refer to 39).
Connect the other end of the extension terminal cable to ZR-RX45. * Press in the cable until it is securely locked. Extension terminal cable Mounting the Extension Terminal Base Set (ZR-XRE1) ZR-RX45 User’s Manual...
Mounting the 20 Channel Extension Terminal Set (ZR-XRT1) Mount the 20 channel extension terminal set as shown below. Important Make sure the ZR-RX45's power is OFF when mounting the extension terminals. Set Contents Extension Terminal Base Unit : 1 20 Channel Terminals : 1 Connection Plate : 1 M4 ×...
M4 × 6 flat head screws Connection plate Connect the extension terminal cable to the extension terminal base unit. * Press in the cable until it is securely locked. Extension terminal cable Mounting the 20 Channel Extension Terminal Set (ZR-XRT1) ZR-RX45 User’s Manual...
Page 47
Connect the other end of the extension terminal cable to ZR-RX45. * Press in the cable until it is securely locked. Extension terminal cable Unit 10 (CH 181 to 200) Unit 2 (CH 21 to 40) Unit 1 (CH 1 to 20) Important When connecting additional terminals, make sure they are added in a continuous manner.
• Maximum input voltage : 60Vp-p • Withstand voltage : 350 Vp-p at 1 minute <Between input terminal/GND (C) > • Maximum input voltage : 60Vp-p • Withstand voltage : 350 Vp-p at 1 minute Precautions to Observe When Performing Measurement ZR-RX45 User’s Manual...
While in the status (open) in which signals are not input to the input terminal, measured results may be influenced by signals from other channels. In such a case, turn OFF the input setting or short circuit +/-. If signals are input correctly, measured results are not influenced by other channels. ZR-RX45 User’s Manual Precautions to Observe When Performing Measurement...
Connect the signal chassis GND and the measurement device chassis ground. Use a short, thick lead to connect the chassis GND of the measurement object to the ZR-RX45' chassis GND. It will become even more effective if the ground potentials are the same.
Setting the Date and Time If you are using the ZR-RX45 for the first time, charge the internal rechargeable battery and then make the date and time settings. Important If the ZR-RX45 is not used for a period of approximately six months, the internal rechargeable battery may be dis- charged and the date and time may revert to the initial settings.
Settings and Measurement This chapter describes the setting and measurement procedures for the ZR-RX45. Window names and functions Key Operation Operation Modes Setting Menus WEB Server Function...
*: Capture stop processing or other cases when data is being written to the built-in internal memory or USB memory : Appears when the ZR-RX45 waits for you to press the Start/Stop key to stop it after data capture. : Appears while the data in the internal memory is being replayed.
: USB memory is being accessed. Do not remove the USB memory. Important Do not remove the USB memory or turn off the power of the ZR-RX45 while the USB memory is being accessed. Data may become corrupt and inaccessible.
: Indicates local mode. Operations can be conducted on the ZR-RX45. : Indicates remote mode. With some exceptions, operations must be conducted on a PC. When you cancel the connection on the application, the ZR-RX45 is automatically sent back to local mode.
13. Pen display Displays the position of each channel signal, trigger and alarm range. Trigger position Alarm range Rising Falling Win In Win Out Stop position Start position ZR-RX45 User’s Manual Window names and functions...
Displays the lower limit scale of the currently active channel. 16. Waveform display Displays the waveform of the input signal. 17. Upper limit scale Displays the upper limit scale of the currently active channel. Window names and functions ZR-RX45 User’s Manual...
* Remaining time more than 99999 hours is displayed as "++++:++:++". During data replay Displays the display position, cursor position, and trigger position graphically. Cursor A position Cursor B position Capture data size Range for which a waveform is currently displayed ZR-RX45 User’s Manual Window names and functions...
6 steps is set. ▪ Change available : 10% steps of range * When ALL is set, setting values for CH1 is reflected on other channels. When CH1 is OFF, ALL Cannot be set. ZR-RX45 User’s Manual Key Operation...
• To return to the MONITOR window when the SPAN/TRACE/POSITION window is displayed. • To cancel remote status (in which keys are disabled) through interface control. • To close the menu screen. • To quit data replay. Key Operation ZR-RX45 User’s Manual...
A password for canceling the key lock can be specified. See page 117 for details. • To change the display mode in the Digital + Calculation Display screen 2ch Mode 4ch Mode 10ch Mode All Mode ZR-RX45 User’s Manual Key Operation...
While pressing the ZR-RX45 START/STOP key, turn the power ON. The external storage media is recognized by the PC and data exchange becomes possible. * In USB Drive Mode, the display on the ZR-RX45 becomes as follows: Important • To exit USB Drive Mode, turn off and on the power again.
• During Free Running, replays captured data. The screen used to specify the data replay source file appears; specify the file you want to replay. • While capturing data, recently captured data is replayed in two windows. ZR-RX45 User’s Manual Key Operation...
Refer to page 81 for details. - Use the FAST FORWARD keys to change the display mode. The calculation results are displayed only in All Mode. Refer to page 61 for details on All Mode. Key Operation ZR-RX45 User’s Manual...
• Saves or reads the currently set condition into the USB device. (can be set during Free Running only). • Exchanges USB memory during data capture (can be set only while data is captured to USB memory). ZR-RX45 User’s Manual Key Operation...
Press the ENTER key to confirm the value. The above explanation shows the basic procedure that may be used for each setting. Please follow the procedure indicated by each menu. Key Operation ZR-RX45 User’s Manual...
File operations The FILE key is used to perform file-related operations. Data replay The REVIEW key is used to replay captured data. Time axis change The TIME/DIV key is used to change the time axis. ZR-RX45 User’s Manual Operation Modes...
While data is replayed in two windows, the FILE key is used to save data to a device. Setting check The MENU key is used to change the settings. Time axis change The TIME/DIV key is used to change the time axis. Operation Modes ZR-RX45 User’s Manual...
While capture to the internal memory is in progress, use this feature to save data to USB memory without stopping capture.) Screen copy The FILE menu is used to copy the screen. ZR-RX45 User’s Manual Operation Modes...
The CURSOR key is used to switch between cursors A and B. keys are used to move the cursors File operations The FILE key is used to save the data between the cursors. Time axis change The TIME/DIV key is used to change the time axis. Operation Modes ZR-RX45 User’s Manual...
Current, length, area, volume, speed, acceleration, frequency, weight, work, pressure, flow rate, temperature Choose (The selections vary depending on the unit selected in the above.) Arbitrary unit Text input Annotation string Text input (Max. 31 characters) ZR-RX45 User’s Manual Setting Menus...
Page 74
(The selections vary depending on the unit selected in the above.) Arbitrary unit Text input Misc Waveform color 0 to 31 for each of red, green, blue (RGB) setting Amplitude Setting 1 to 8 dots Setting Menus ZR-RX45 User’s Manual...
Page 75
Input item Description Voltage 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 1-5 V Temperature TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N, TC-W, Pt100, JPt100, Pt1000 Humidity No selection available ZR-RX45 User’s Manual Setting Menus...
Page 76
50°C -200 to +850°C JPt100 -270 to +2000°C 50°C -200 to +500°C Pt1000 -270 to +2000°C 50°C -200 to +500°C <Humidity Range> Range Maximum SPAN Minimum SPAN (p-p) Minimum Resolution 0 to +110% 1.0% 0.1% Setting Menus ZR-RX45 User’s Manual...
Page 77
A moving average is calculated 40 times per sampling interval. <Filter processing> Filter processing performed on the ZR-RX45 is the moving average shown in the following figure. If the filter setting is 5, D=(n-4 + n-3 + n-1 + n)÷5...
Page 78
• The Scaling function performs calculation using a ratio of the Meas. Value and EU Output Value settings. The dig- ital display shows "++++/----" when the converted value cannot be processed by the ZR-RX45. • The span may be changed depending on the Scaling settings.
Page 79
* For temperature input, this function is not available. (6) Reset Auto Zero ADJ. Cancels the zero position voltage value and displays the input voltage. (7) [Zero position voltage Displays the zero position voltage value (Display only). value] ZR-RX45 User’s Manual Setting Menus...
Page 80
When "Select" or "Choose" is used, the setting is reflected in this field. N o te • Calculation results are displayed in volts. The calculation result for 100 mV + 100 mV is 0.2. Use Scaling to display this result as 200 mV. Setting Menus ZR-RX45 User’s Manual...
Page 81
Captures the number of pulses for each sampling interval. (1)-9 Filter Sets the filter for input. Selection item Description Disables hardware filter. Enables hardware filter. It is effective in a noisy environment. The filter is approximately 30 Hz (-3 dB). ZR-RX45 User’s Manual Setting Menus...
Page 82
• If a message window opens, follow the instruction in the message to change the setting value. • The Scaling function performs calculation using a ratio of the Meas. Value and EU Output Value settings. The digital display shows "++++/----" when the converted value cannot be processed by the ZR-RX45. Setting example Meas.
Page 84
Specify a folder to which you want to capture (or save) data. Refer to page 109, "File box" for details. (2) File Specify a file to which you want to capture (or save) data. Refer to page 109, "File box" for details. Setting Menus ZR-RX45 User’s Manual...
Page 85
Example: TEST_CP1.GBD (4) File Type Sets the file format used to save data. GBD : Creating a data file in OMRON's proprietary binary format * Data tampering can be prevented. CSV : Creating a data file in text format * Replaying on the ZR-RX45 is not available.
Page 86
When capturing is stopped at the STOP poinf in the above, the File4 and the File 5 are remained. These files are consolidated into one file and it is saved. Then the ring catpuer is completed. File4 File5 Filexx : Saved ••• Setting Menus ZR-RX45 User’s Manual...
Page 87
Enable this setting to enable the digital filter. - When you use external sampling and measure signals with high noise levels, set the AC line filter to ON. The shortest interval is displayed under the settings. Shortest interval ZR-RX45 User’s Manual Setting Menus...
Page 88
The Number of Measuring Channels is the number of channels for which the input setting is not OFF. (2)-6 Backup setting The ZR-RX45 has a function that periodically backs up captured data (See the figure below). This section explains how to set the data backup conditions.
Page 89
All Mode. Refer to page 61 for details on All Mode. • At power-on, calculation is started; The result is cleared when you press the QUIT key or the Start key to start measurement. ZR-RX45 User’s Manual Setting Menus...
From 2005.1.1 to 2035.12.31 Time From 0:0:0 to 23:59:59 [Weekly] Day of week Off or On setting for each of Sunday through Saturday Time From 0:0:0 to 23:59:59 [Time] From 0:0:1 to 9999:59:59 Repeated Capturing Off, On Setting Menus ZR-RX45 User’s Manual...
Page 91
Example: On is set for Mon, Tue, Wed, Thu, and Fri, Off is set for Sun and Sat, and 9:00 is set as the time. Starts capturing data at 9:00 on weekdays. Does not start capturing data on Sat and Sun. Time Starts capturing data when a specified length of time elapses. ZR-RX45 User’s Manual Setting Menus...
Page 92
The repeat function is disabled. The repeat function is enabled. After one capture is ended, the next capture is started (If the start side source setting is not Off, the ZR-RX45 waits for a trigger). (3)-4 Alarm level settings Sets alarm generation conditions, output destination, etc.
Page 93
<Trigger level settings> <Alarm level settings> Place the cursor here Place the cursor here and press the ENTER and press the ENTER key to open the key to open the following setting screen. following setting screen. ZR-RX45 User’s Manual Setting Menus...
Page 94
An edge exceeding the setting level is required. For 2CH, a trigger is generated here. Measurement is Measurement is Measrement is started here. immediately started from here in started in the the Edge operation. Level operation. Setting Menus ZR-RX45 User’s Manual...
Page 95
: Used to specify the upper and lower limits for each channel. When the input signal level goes (or is) out of these limits, a trigger/alarm is generated. Trigger (alarm) generation area Measurement starts (alarm generated) Upper limit level Lower limit level CH.1 ZR-RX45 User’s Manual Setting Menus...
Page 96
Setting value + 0.5% of range (For 1 V range, 1 V*0.5%=0.005 V) Trigger/alarm setting level Setting value - 0.5% of range (For 1 V range, 1 V*0.5%=0.005 V) Conditions met here in rising Conditions met here in falling Setting Menus ZR-RX45 User’s Manual...
Specify the department name. You cannot specify it as Guest. Setting conditions switch Switches between Guest, User1 and User2. Since setting conditions are stored for each user, they can be called up easily by simply switching the user. ZR-RX45 User’s Manual Setting Menus...
(5)-1 New Line code Specifies the line feed code. Selection item Description CR+LF Starts a new line with CR+LF code (default value). Starts a new line with LF code. Starts a new line with CR code. Setting Menus ZR-RX45 User’s Manual...
Page 99
Sets the USB ID number of ZR-RX45. Specify a number from 0 to 9 (default value: 0). To control more than one ZR-RX45 unit with one PC, assign a unique USB ID to each of them. (5)-3 TCP-IP settings TCP-IP settings are used to connect the ZR-RX45 to Ethernet.
Page 100
Performs connection test to the FTP server.When the connection test is performed, Test a message is displayed.If connection cannot be established, check the settings and perform the connection test again. * If the connection test is passed, the following message is displayed. Setting Menus ZR-RX45 User’s Manual...
Japanese, English (US), English (UK), French, German, Chinese, Korean Return to default settings Press right key to execute. Information Down button to display information Demo Waveform Mode Off, On (6)-1 LCD brightness Sets the brightness of the LCD backlight. ZR-RX45 User’s Manual Setting Menus...
Page 102
Periodical burnout check is conducted. Important During a burnout check, voltage is applied to the ZR-RX45. Therefore, set Burn Out to "Off" when ZR-RX45 is con- nected in parallel with other devices to avoid any effect from these voltages. Setting Menus...
Page 103
(6)-9 Date/Time Makes settings related to the ZR-RX45 clock. The internal clock (date and time) of the ZR-RX45 can be set. Alternatively, if the Network Time setting is used, the ZR-RX45 clock can be automatically adjusted via the network. Refer to the next section, "Network Time Setting" for details.
Page 104
Network Time setting The ZR-RX45 has a function that synchronizes the time to that of a time server via Ethernet. This section describes settings to be made to use this function. Setting Selections available Network Time Enables or disables this function.
Refer to page 67 for details on the operation modes. <Free Running Status> <Replay or Double-Screen Replay Status> <Capture Status> (7)-1 File Operation Operate files in the main memory and USB device. For details on file operation, see on page 109. ZR-RX45 User’s Manual Setting Menus...
Page 106
Specify a file to which you want to save data. Refer to page 109, "File box" for details. (3) File Type Sets the file format used to save data. GBD: Creating a data file in OMRON's proprietary binary format * Data tampering can be prevented. CSV: Creating a data file in text format * Replaying on the ZR-RX45 is not available.
Page 107
(7)-3 Remove/Exchange USB Memory The ZR-RX45 allows you to exchange the USB memory while data is captured to it. Exchange the USB memory in accordance with the following procedure: Press the FILE key to open the FILE menu. Move the cursor to Remove/Exchange USB Memory and press the ENTER key.
Page 108
Third USB memory device : TEST_CHG2.GBD * While ring capture is On, the USB memory device cannot be exchanged. Important The exchange procedure must be completed within ten minutes. Data will be lost after ten minutes have elapsed. Setting Menus ZR-RX45 User’s Manual...
Page 109
PNG: Saves data in ping format (7)-5 Execute (Screen Copy) Executes screen copy and saves it to an image file. * Refer to page (7)-4 "Specify Save Destination" for details on specifying the save destination. ZR-RX45 User’s Manual Setting Menus...
Page 110
: Data is captured to a file with an entered file name. Sequential number : A file is created with an arbitrary file name that has been entered, followed by a sequential number. (7)-7 Load Loads and reflects the ZR-RX45 setting conditions from a file. Setting Description (1) File Specify a file to which you want to save data.
Close the file box. <Setting example> The following shows an operation example where a folder named "TEST" is created for captured data and automatically saved. In the [Data save Destination], choose [Select folder] and press the ENTER key. ZR-RX45 User’s Manual Setting Menus...
Page 112
When the input box for a new folder name opens, enter "TEST" and click OK. Use the key to choose [Select file/folder]. Use the key to move the cursor to the created "TEST" folder, and press the ENTER key. Setting Menus ZR-RX45 User’s Manual...
When selecting operation Text used for each operation When you bring the cursor to a character and press ENTER, the character is entered. After you finish entering characters, move the cursor to OK and then press ENTER. ZR-RX45 User’s Manual Setting Menus...
Set numeric value Set numeric value Find Next Press right key to execute. Find Previous Press right key to execute. Statistical calculation Function Off, Average, Max, Min, Peak, RMS between cursors Execute Press right key to execute. Setting Menus ZR-RX45 User’s Manual...
Page 115
Cursors are not synchronized. Only the specified one cursor moves. Two cursors move in synchronization. Cursor A is always the fulcrum. Cursor Synch is turned Off when you move a cursor using Move to Selected Position or perform Data Search. ZR-RX45 User’s Manual Setting Menus...
Page 116
(10)-8 Find Previous Executing this option moves the cursor to a previous position where the search conditions are met, up from the current cursor position. (Specify the search conditions as described in (10)-6 "Data Search.") Setting Menus ZR-RX45 User’s Manual...
Page 117
Displays the peak value of the data during data capture. Displays the RMS value of the data during data capture. The calculation formula is as fol- lows: R.M.S = * D: data n: number of data ZR-RX45 User’s Manual Setting Menus...
Replaying SERCH key can be used to perform search. : Searches past side : Searches future side ZONE key can be used to change the zone division. Setting Menus ZR-RX45 User’s Manual...
(13) To cancel key lock by password A password can be set to ZR-RX45 to cancel the key lock. (No password is set at factory default.) <Operation flow> Set the password. Press the , and ENTER keys at the same time to display the password setting screen shown below.
• http ....Protocol to access the server. HTTP (Hyper Text Transfer Protocol) • IP address ..Type in the IP address of the ZR-RX45 to monitor. • Index.html..File name. This is fixed to Index.html. N o te The port number can be omitted.
Page 121
Zoom........ Enlarges only the LCD screen of ZR-RX45. Digital ....... Displays the ZR-RX45 measured value digitally. Download of device file ..Allows data captured with ZR-RX45 to be downloaded to your PC via FTP. Remote key operation To operate ZR-RX45 from a remote location, click the corresponding ZR-RX45's panel keys on the screen.
Page 122
Screen update speed..... Specifies the speed in which the screen is updated. Available update speeds are 2, 5, and 10 seconds. Digital Screen update speed..Specifies the speed in which the screen is updated. Available update speeds are 2, 5, and 10 seconds. WEB Server Function ZR-RX45 User’s Manual...
Page 123
• Upload file • Delete file/folder • Create file/folder • Change file name/folder name To enable data to be written to the ZR-RX45, the login account name must be changed. please use the following table as a guide. Account name Password...
Page 124
Select [Tools] - [Internet Options] to open Internet Options. Click the [Advanced] tab and check Enable FTP Folder View (Outside of Internet Explorer). Click the OK button to close the Internet Options dialog. Close Internet Explorer. Open Explorer and re-enter the URL. WEB Server Function ZR-RX45 User’s Manual...
Page 125
Finally, click the Logon button. N o te If Explorer does not display the File menu, enable the display of this menu by selecting [Organize] - [Layout] - [Menu Bar] and checking this option. ZR-RX45 User’s Manual WEB Server Function...
Page 126
MEMO WEB Server Function ZR-RX45 User’s Manual...
When the screensaver is operating 0.65 A Recharging bat- tery is not possi- Vibration resistance Equivalent to automobile parts Type 1 Category A classification External dimensions 232 × 152 × 50 mm SPECIFICATIONS ZR-RX45 User’s Manual...
Page 129
*3 *4 temperature (°C) Accuracy detector Pt100 -200 to 850 (FS = 1050°C) 1 mA ±1.0°C JPt100 -200 to 500 (FS = 700°C) 1 mA ±0.8°C Pt1000 -200 to 500 (FS = 700°C) 0.2 mA ±0.8°C ZR-RX45 User’s Manual SPECIFICATIONS...
Page 130
<Maximum ratings of output transistor> • Collector-GND voltage : 30 V • Collector current : 0.5 A • Collector dissipation : 0.2 W Output conditions Level judgment, window judgment, logic pattern judgment, pulse judgment SPECIFICATIONS ZR-RX45 User’s Manual...
Page 131
PC Software connections: Transfers data to the PC (realtime, memory), controls of (10BASE-T/100BASE-TX) ZR-RX45. Web server function: Displays ZR-RX45's screen image on Web browser, operates of ZR-RX45. FTP server function: Transfers and deletes files from internal memory and USB memory.
Method: Realtime and between cursors specified (during data replay) Calculation results are displayed in Digital screen + Calculation Display screen. Calculation between Calculation type: Addition, subtraction, multiplication, and division channels Input channels: Analog channels 1 through 200 SPECIFICATIONS ZR-RX45 User’s Manual...
Page 133
A comment can be input for each channel. function Inputtable characters : Alphanumerics Number of characters : 31 The settings of 50 ms or below can be used depending on the input settings and the measuring channels. ZR-RX45 User’s Manual SPECIFICATIONS...
PC Software Item Special PC software Standard PC software Wave Inspire RX (Ver. 2.4) Smart Viewer RXW (Ver. 2.0) Compatible logger ZR-RX70, ZR-RX45, ZR-RX25, ZR-RX45, ZR-RX25, ZR-RX40, ZR-RX40, ZR-RX20 ZR-RX20 Compatible operating Windows XP / Vista / 7 system Intel (x86) compatible processor 1 GHz or Pentium 4: 2.0GHz or faster professor...
When capturing to internal memory at 1s sampling, 20 channel terminals, using new battery packs at +25°C environment. (The running time depends on the operating environment.) Charging method Mount in the main unit (ZR-RX45) Time required for charging Battery pack × 1: approx. 4 hours Battery pack × 2: approx. 8 hours...
Page 136
Item Description Sensor output 0 to 1 VDC External dimensions 14 mm × 80 mm (excluding cable) Cable length Sensor power source 5 to 16 VDC Power consumption Approx. 4 mA Accessory/Option Specifications ZR-RX45 User’s Manual...
Invalid trigger stop settings. When logging is started with the trig- Please set the stop trigger correctly. Press [Enter] key and change the set- ger setting not normally set,it is dis- tings. played.(STOP) Error Messages and Countermeasures ZR-RX45 User’s Manual...
"Disk is full" is usually displayed. cause. Please increase the capacity of the • Disk is full. disk as it erases an unnecessary • Neither the file nor the directory file. are created any further. ZR-RX45 User’s Manual Error Messages and Countermeasures...
Page 142
MEMO Error Messages and Countermeasures ZR-RX45 User’s Manual...
Clock display area FTP server setting 97, 98 Connecting to a PC Function Specifications Connection Diagram CURSOR key Cursor Position Gateway Cursor Sync Guest Data Capture bar Humidity sensor 39, 133 Data replay menu Data Save ZR-RX45 User’s Manual Index...
Page 144
Statistical calculation setting NAVI menu Status mark area Network Time Stop side source setting New Line code Subnet Mask Noise Countermeasures Synchronization Mode NTP Server Synchronization Time Operating Environment TCP-IP settings OTHER setting Temp Unit Terminal unit Index ZR-RX45 User’s Manual...
Page 145
USB Drive Mode USB ID USB Memory access display area Attaching Remove/Exchange USB settings User USER settings Warming-up Waveform color setting 77, 81 Waveform display area WEB Server Window names and functions Zero Adjustment Zero position voltage value ZR-RX45 User’s Manual Index...
A manual revision code appears as a suffix to the catalog number at the bottom of the front and back covers of this manual. Cat. No. Z304-E1-01 Revision code Revision code Date Revised contents September 2010 Original production Revision History ZR-RX45 User’s Manual...