Portable Multi Logger
ZR-RX40
User's Manual
Cat. No. Z260-E1-04

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ZR-RX40 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Omron ZR-RX40

  • Page 1 Portable Multi Logger ZR-RX40 User's Manual Cat. No. Z260-E1-04...
  • Page 2 · Before use: connection and wiring in details, language change of display, etc. · Procedure in details for setting and measurement · Specifications of the ZR-RX40 series and accessories · Other information which is required for the use of the ZR-RX40 series...
  • Page 3: Application Considerations

    APPLICATION CONSIDERATIONS (Please Read) BEFORE USE User's Manual SETTINGS AND MEASUREMENT SPECIFICATIONS APPENDIX Portable Multi Logger ZR-RX40...
  • Page 4 LOSS OF PROFITS OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR STRICT LIABILITY. In no event shall responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.
  • Page 5 Performance data given in this document is provided as a guide for the user in determining suitability and does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users must correlate it to actual application requirements. Actual performance is subject to the OMRON Warranty and Limitations of Liability.
  • Page 6 Indicates the possibility of electric shock under specific conditions. Indicates prohibition when there is a risk of minor injury from electrical shock or other source if the product is disassembled. Indicates general prohibitions for which there is no specific symbol. ZR-RX40 User’s Manual...
  • Page 7 Hazard may occur from serious fire or electric shock. Do not connect voltages exceeding the rated voltage to the signal input terminals. Fire or hazard may occur in rare occasions from ignition, rupture or combustion. Do not use battery packs other than ZR-XRB1. ZR-RX40 User’s Manual...
  • Page 8 II, III and IV. 4.Others • Dispose of this product as industrial waste. • If there are any troubles, stop usage immediately, turn off the power supply and contact OMRON branch or sales office. ZR-RX40 User’s Manual...
  • Page 9 • Do not drop or apply strong impact or force to the product. It may cause malfunction of the monitor or the main unit. 6.Maintenance • Do not use thinner, benzine, acetone or kerosene to clean this product. • Calibration should be performed periodically to maintain measurement accuracy. ZR-RX40 User’s Manual...
  • Page 10 • Specal PC software “Wave Inspire RX“ (tryout) • Basic PC software “Smart Viewer RXW” • User's Manual PDF files (this manual) • “Wave Inspire RX“ Software Manual PDF files • “Smart Viewer RXW” Software Manual PDF files ZR-RX40 User’s Manual...
  • Page 11 Editor's Note Meaning of Symbols Menu items that are displayed on the ZR-RX40's LCD screen, and windows, dialog boxes and other GUI elements displayed on the PC are indicated enclosed by brackets "[ ]". Visual Aids Indicates points that are important to achieve the full product performance, Important such as operational precautions.
  • Page 12 MEMO ZR-RX40 User’s Manual...
  • Page 13: Table Of Contents

    Changing the Display Language ....... . 40 ZR-RX40 User’s Manual...
  • Page 14 External Dimensions......... 100 ZR-RX40 User’s Manual...
  • Page 15 Revision History ..........108 ZR-RX40 User’s Manual...
  • Page 16 ZR-RX40 User’s Manual...
  • Page 17 BEFORE USE Features Part Names and Functions Connecting the Power Cable Connecting the Analog Input Terminal Connecting the Logic Input Terminal Attaching USB Memory Device Connecting to a PC Using the Battery Pack Connecting the Humidity Sensor Mounting and Removing the Terminal Unit Mounting the Extension Terminal Base Set Mounting the 20 Channel Extension Terminal Set Precautions to Observe When Performing Measurement...
  • Page 18: Features

    • The USB drive mode function enables the ZR-RX40's internal memory to be recognized as an external drive by your PC. (Connect the ZR-RX40 to your PC and turn on the power supply to the ZR-RX40 while holding down the [START] key.)
  • Page 19: Part Names And Functions

    Part Names and Functions This section describes the names and function of parts of the ZR-RX40. Power switch PC interface terminals Monitor Operation status LED POWER : ON when the power is ON START : ON during data capture CHARGE...
  • Page 20: Connecting The Power Cable

    Be sure to use only the AC cable and the AC adapter provided as standard accessories. Plug the AC cable into the AC adapter. AC Aadapter AC Cable Connect the output side of the AC adapter to the AC adapter connector. AC adapter cable Connecting the Power Cable ZR-RX40 User’s Manual...
  • Page 21: Connecting To A Dc Power Supply

    Press the power switch on the ZR-RX40 to the ON side to turn on the power. Important Connect the GND terminal for safe measurement. The ZR-RX40 must also be grounded when sharing a common ground level with other devices. Connecting to a DC Power Supply Use the DC cable (option: ZR-XRD1).
  • Page 22 Connect the DC input side to the DC power supply. Important Be sure to check the polarity of the power supply when connecting the DC cable. Press the power switch on the ZR-RX40 to the ON side to turn on the power. Connecting the Power Cable ZR-RX40 User’s Manual...
  • Page 23: Connecting The Analog Input Terminal

    20, 50, 100, 200, 500 mV/F.S.; 1, 2, 5, 10, 20, 50 V/F.S.; 1-5V Thermocouples K, J, E, T, R, S, B, N, W (WRe 5-26) Resistance temperature detector PT100, JPT100, PT1000 (IEC751) Connecting the Analog Input Terminal ZR-RX40 User’s Manual...
  • Page 24: Connecting The Logic Input Terminal

    Internal Equivalence Circuit and Example of Wiring Connection Item Description Number of output channels 4 Output format Open collector output +5 V, 10 KΩ pull-up resistance Contact capacity 5 V to 24 V, 100 mA or below Connecting the Logic Input Terminal ZR-RX40 User’s Manual...
  • Page 25 White with red dotted line White with black dotted line Alarm output Yellow with red dotted line Yellow with black dotted line Pink with red dotted line : Trigger input Pink with black dotted line Shielded Connecting the Logic Input Terminal ZR-RX40 User’s Manual...
  • Page 26: Attaching Usb Memory Device

    Attaching USB Memory Device Attaching USB memory device to the ZR-RX40 allows you store measured data directly. Inserting a USB Memory Device Attach the USB memory device to the USB memory terminal. USB memory device Important <Specifications of supported USB memory>...
  • Page 27: Connecting To A Pc

    Connecting to a PC Use the USB, LAN Interface to connect the ZR-RX40 to a PC. Connection Using a USB Cable Use the USB cable to connect the ZR-RX40 to a PC. USB cable Note If the USB cable is used, install the USB driver in your PC.
  • Page 28: Lan Connection

    LAN Connection Use a LAN cable to connect the ZR-RX40 to a PC. LAN cable Cable Types • Use a crossing cable when connecting directly to a PC, without using a hub. LAN cable (crossing) • Use a straight cable when using a hub.
  • Page 29: Using The Battery Pack

    Do not use a new battery with an old battery at the same time. When attaching two battery packs, make sure the remaining amount are same. If you are not sure about the amount, charge each battery and then attach full-charged two battery packs. Using the Battery Pack ZR-RX40 User’s Manual...
  • Page 30: Charging The Battery

    • battery pack x 1: approx. 4 hours • battery pack x 2: approx. 8 hours The battery pack is charged by mounting it in the ZR-RX40, attaching AC adapter to the ZR-RX40. Mount the battery pack in the ZR-RX40.
  • Page 31: Connecting The Humidity Sensor

    Connecting the Humidity Sensor Connect the + and - lead wires of the humidity sensor (option: ZR-XRH1) to the desired terminals, and then insert the round connector into the 5V OUT connector on the ZR-RX40. Connected to the 5V OUT terminal.
  • Page 32: Mounting And Removing The Terminal Unit

    Remove and mount terminal units as shown below. Important Make sure the ZR-RX40’s power is OFF when removing or mounting terminal units. To Remove Pull the terminal unit out towards the direction indicated by the arrow while pressing the two locks at the bottom of the unit.
  • Page 33: To Mount

    To Mount Insert the tabs at the top of the terminal unit into the slots of the ZR-RX40, and push in the unit until the lock tabs at the bottom of the unit are securely locked. (1) Insert tabs at the top of the terminal unit into the slots.
  • Page 34: Mounting The Extension Terminal Base Set

    Mounting the Extension Terminal Base Set Mount the extension terminal base set (ZR-XRE1) as shown below. Important Make sure the ZR-RX40’s power is OFF when mounting the extension terminals. Set Contents Extension Terminal Base Unit : 1 Extension Terminal Cable : 1...
  • Page 35 Connect the other end of the extension terminal cable to ZR-RX40. * Press in the cable until it is securely locked. Extension terminal cable Mounting the Extension Terminal Base Set ZR-RX40 User’s Manual...
  • Page 36: Mounting The 20 Channel Extension Terminal Set

    Mounting the 20 Channel Extension Terminal Set Mount the 20 channel extension terminal set (ZR-XRT1) as shown below. Important Make sure the ZR-RX40’s power is OFF when mounting the extension terminals. Set Contents Extension Terminal Base Unit : 1 20 Channel Terminals : 1...
  • Page 37 M4 x 6 flat head screws Connection plate Connect the extension terminal cable to the extension terminal base unit. * Press in the cable until it is securely locked. Extension terminal cable Mounting the 20 Channel Extension Terminal Set ZR-RX40 User’s Manual...
  • Page 38 Connect the other end of the extension terminal cable to ZR-RX40. * Press in the cable until it is securely locked. Extension terminal cable Terminal unit 10 (CH 181 to CH 200) Terminal unit 2 (CH 21 to CH 40)
  • Page 39: Precautions To Observe When Performing Measurement

    Before starting another measurement operation, short-circuit the + and - terminals to enable self- discharge. The ZR-RX40 has a scan system. While in the status (open) in which signals are not input to the input terminal, measured results may be influenced by signals from other channels.
  • Page 40: Noise Countermeasures

    Connect the signal chassis GND and the ZR-RX40’s chassis GND. Use a short, thick lead to connect the chassis GND of the measurement object to the ZR-RX40’s chassis GND. It will become even more effective if the ground potentials are the same.
  • Page 41: Setting The Date And Time

    Setting the Date and Time The ZR-RX40 includes a rechargeable internal battery for backup. If you are using the ZR-RX40 for the first time, charge the internal rechargeable battery and then make the date and time settings. Note If the ZR-RX40 is not used for a period of approximately six months, the internal rechargeable battery may be discharged and the date and time may revert to the initial settings.
  • Page 42: Changing The Display Language

    Changing the Display Language You can choose the language displayed on the screen. The default display language is set to English when the ZR-RX40 is shipped overseas. ZR-RX40A: Japanese ZR-RX40A-E,-U,-B,-CHRO: English To change the display language, see the instructions in "OTHR:Language".
  • Page 43: Settings And Measurement

    SETTINGS AND MEASUREMENT Window names and functions Key Operation Operation Modes Setting Menus WEB Server Function...
  • Page 44: Window Names And Functions

    4. Key lock display: Displays the key lock status. (Yellow = Locked) 5. Remote display: Lit when the ZR-RX40 is in remote mode. (Yellow = Remote mode) 6. Clock display: Displays the current date and time. 7. AC/battery status display: Displays the icon when AC power is used and indicates the level when the battery is used.
  • Page 45 Displays the upper limit scale of the currently active channel. 17. Data capture bar: Indicates the remaining capacity of the capture media during data capture. During replay, indicates information about the displayed position. Window names and functions ZR-RX40 User’s Manual...
  • Page 46: Key Operation

    • Switch channels of the AMP settings • Switch channels of the trigger/alarm level settings • Switch channels of the calculation display 1 to 10 11 to 20 191 to 200 MONITOR AMP settings Level settings Digital screen Key Operation ZR-RX40 User’s Manual...
  • Page 47: Span/Trace/Position

    When ALL is set, setting values for CH1 is reflected on other channels. Note * When CH1 is OFF, ALL cannot be set. Key Operation ZR-RX40 User’s Manual...
  • Page 48: Time/Div

    • To cancel remote status (in which keys are disabled) through interface control. • To close the menu screen. • To quit data replay. • To return the Enlarged Waveform Screen/Digital + Calculation Display Screen to the Waveform + Digital Screen. Key Operation ZR-RX40 User’s Manual...
  • Page 49: Direction Keys

    • To set key lock. (Hold down the left/right FAST FORWARD key for at least two seconds. Press again to unlock.) A password for canceling the key lock can be specified. "To cancel key lock by password" p.48 Key Operation ZR-RX40 User’s Manual...
  • Page 50: Start/Stop (Usb Drive Mode)

    In USB Drive Mode, the internal memory is recognized by the PC as external storage media. USB Drive Mode Operation Procedure Use a USB cable to connect the ZR-RX40 and a PC. (When the USB driver has not been installed, install it as described in the Software Manual "Installing the USB Driver".) Connect the USB device to the ZR-RX40.
  • Page 51: Review

    • During Free Running, replays captured data. The screen used to specify the data replay source file appears; specify the file you want to replay. • While capturing data, recently captured data is replayed in two windows. Key Operation ZR-RX40 User’s Manual...
  • Page 52: Display

    <Waveform + Digital Screen> Default <Expanded Waveform screen> Displays the waveform in full screen mode. <Digital + Calculation Display screen> Displays digital values and calculation results in large text. * Calculation results are added up in real time. Key Operation ZR-RX40 User’s Manual...
  • Page 53: Cursor (Alarm Clear)

    • Saves data between cursor A and cursor B during replay (can be set during replay only). • Saves the settings currently in use or loads settings. • Saves or reads the currently set condition into the USB device. Key Operation ZR-RX40 User’s Manual...
  • Page 54: Navi

    Press the [ENTER] key to confirm the value. The above operation is the basic procedure that may be used for each setting. However, precise procedure may vary between setting items. Please follow the procedure indicated by each menu. Key Operation ZR-RX40 User’s Manual...
  • Page 55: Operation Modes

    The [SPAN/TRACE/POSITION] key is used to change settings. Display mode The [DISPLAY] key is used to change the display mode. File operations The [FILE] key is used to perform file-related operations. Data replay The [REVIEW] key is used to replay captured data. Operation Modes ZR-RX40 User’s Manual...
  • Page 56: Capturing

    The [CURSOR] key is used to switch between cursors A and B. The left/right or FAST FORWARD keys are used to move the cursors. Important Captured data can only be displayed for the double window buffer portion. Dual view buffer capacity is 512 KB. Operation Modes ZR-RX40 User’s Manual...
  • Page 57: Replaying

    The [CURSOR] key is used to switch between cursors A and B. The left/right or FAST FORWARD keys are used to move the cursors File operations The [FILE] key is used to save the data between the cursors. Operation Modes ZR-RX40 User’s Manual...
  • Page 58: Setting Menus

    Unit Meas. Value (Upper/Lower) EU Value (Upper/Lower) Dec pt Unit Select Misc. Span setting (Span All Settings) Annotation setting Zero voltage adjustment Perform Auto Zero ADJ. Reset Auto Zero ADJ. [Zero point voltage value] Setting Menus ZR-RX40 User’s Manual...
  • Page 59 Range ........Specifies the range of signal input to be measured. Voltage : 20, 50, 100, 200, 500mV, 1, 2, 5, 10, 20, 50, 1-5V Temperature : TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N, TC-W, PT100, JPT100, PT1000 Setting Menus ZR-RX40 User’s Manual...
  • Page 60 50°C –200 to +850°C JPT100 –270 to +2000°C 50°C –200 to +500°C PT1000 –270 to +2000°C 50°C –200 to +500°C <Humidity Range> Range Maximum SPAN Minimum SPAN (p-p) Minimum Resolution 0 to +110% 1.0% 0.1% Setting Menus ZR-RX40 User’s Manual...
  • Page 61 The Scaling operation is calculated using a ratio of the Meas. Value or EU Output Value settings. "++++/----" is displayed when the converted value cannot be processed by ZR-RX40. Meas. Value Specifies the numeric value to be converted. Set two points, the Upper and Lower parameters.
  • Page 62 60 as rpm values. • Counts : Displays the cumulative number of pulses for each sampling interval from the start of measurement. • Inst. : Displays the number of pulses for each sampling interval. Setting Menus ZR-RX40 User’s Manual...
  • Page 63: Data Settings

    Record Settings 100, 200, 500ms 1, 2, 5, 10, 20, 30s • Sampling Interval 1, 2, 5, 10, 20, 30min, 1h File Name File: Folder name, file name Name Type: Auto, User File Type: GBD, CSV Setting Menus ZR-RX40 User’s Manual...
  • Page 64 : Captures data using a user-defined name. File Format......Set the file format in which data is saved. • GBD : Binary format • CSV : CSV file format (such data cannot be replayed with the ZR-RX40) Setting Menus ZR-RX40 User’s Manual...
  • Page 65 Backup Settings ..... ZR-RX40 provides a back up function where captured data can be backed up periodically (see figure below). This section allows you to set data back up conditions. • Backup Intervals: Sets the interval to perform back up of captured data. Off, 1, 2, 6, 12, 24 h •...
  • Page 66: Trig Settings

    Off). : Starts capturing data when a specified time arrives (when repeated capturing is set to On). This setting is used when you want to start capturing data at the same time every day. Setting Menus ZR-RX40 User’s Manual...
  • Page 67 • Does not occur: Alarm signal is not output to the alarm output terminal when burnout occurs. • Occurs: Alarm signal is output to the alarm output terminal when burnout occurs. Alarm is output to the output terminal set in the Alarm level settings. Setting Menus ZR-RX40 User’s Manual...
  • Page 68 Measurement starts (alarm generated) CH.1 Win In: Used to specify the upper and lower limits for each channel. When the signal level goes within (or is within) either limit, a trigger/alarm is generated. Measurement starts (alarm generated) CH.1 Setting Menus ZR-RX40 User’s Manual...
  • Page 69 Combination (for trigger only)..Sets the combination of trigger conditions set for each channel. Starts (stops) capturing data when at least one trigger condition is met. AND: Starts (stops) capturing data only when all trigger conditions are met. Setting Menus ZR-RX40 User’s Manual...
  • Page 70 These conditions can be set when they have been enabled in the AMP settings. • Off: No trigger/alarm conditions set. • H: Enabled when the logic signal goes from Low to High. • L: Enabled when the logic signal goes from High to Low. Setting Menus ZR-RX40 User’s Manual...
  • Page 71: User Settings

    • Macro file name: Specify the macro file name to be executed. • Macro run: Executes the specified macro file. Setting Menus ZR-RX40 User’s Manual...
  • Page 72: Interface Settings

    : Starts a new line with CR code. USB........Specify settings relative to USB interface. • USB ID : Sets the USB ID number of ZR-RX40. Specify a number from 0 to 9 (default value: 0). TCP-IP ........Specify TCP-IP settings of ZR-RX40.
  • Page 73 You must restart ZR-RX40 when any change is made to an interface setting value. If you continue to operate ZR-RX40 without restarting after these changes, the new setting values will not be applied. As a result, the ZR-RX40 will not be able to establish a connection to your PC.
  • Page 74: Othr Settings

    Turning off the display frequently using the Screen Saver function allows longer operational time with batteries and longer lifetime of the LCD screen. Power On Start ......Sets the feature which initiates measurement as soon as the ZR-RX40 is turned on. • Disable : Disables the Power On Start function.
  • Page 75 • On: Enables periodic burnout check. Important During a burnout check, voltage is applied to the ZR-RX40. Therefore, set Burn Out to "Off" when ZR-RX40 is connected in parallel with other devices to avoid any effect from these voltages. AC Line cycle......Select the frequency of the AC line used.
  • Page 76 • Synchronized Time: Sets the time when ZR-RX40 is synchronized with the time server. The internal clock will start synchronization upon the set time using the method selected in Adjust Mode. • Adjust Mode : Sets the method to synchronize the internal clock to the time server.
  • Page 77: Other Menus

    User. Name Type: Specifies how files are named. • Auto : Automatically uses the capture start time as the file name. • User : Sets to a user-defined name. Execute: Executes bitmap save. Setting Menus ZR-RX40 User’s Manual...
  • Page 78 Specifies the file format used to save data. • GBD : Binary format • CSV : CSV file format (such data cannot be replayed with the ZR-RX40) Name Type: Specifies how to name a file. • Auto : Automatically uses the capture start time as the file name.
  • Page 79 <Operation example when USB thumb drives are exchanged detaching> The following message is displayed. Remove the USB memory device from the ZR-RX40. Insert a new USB memory device. Press the [ENTER] key. Note "_CHG" and a number will be appended to the file name each time you exchange a USB memory device.
  • Page 80 Format disk ........Format the disk. * Details of allowed operation will depend on the operation target. Moves between folders. Move up one folder. : Move down one folder. ENTER Finalize the operation. QUIT Close the file box. Setting Menus ZR-RX40 User’s Manual...
  • Page 81 [ENTER] key. Use the key to move to the target folder. Use the key to select [Create new folder]. Press the [ENTER] key. In the [New folder name] box that appears, type in "TEST". Setting Menus ZR-RX40 User’s Manual...
  • Page 82 Use the key to choose [Select file/folder]. Use the key to move the cursor to the created "TEST" folder, and press the [ENTER] key. Select [OK] to close the screen. Setting Menus ZR-RX40 User’s Manual...
  • Page 83 Text used for each operation When you bring the cursor to a character and press operation ENTER, the character is entered. After you finish entering characters, move the cursor to OK and then press ENTER. Setting Menus ZR-RX40 User’s Manual...
  • Page 84 : Both, H, L • Level : Level: set numeric value (selectable only when analog or pulse) • Next Search • Prev. Search Statistical Calculation between • Statistical Calculation: Off, Average, Max, Min, Peak, RMS cursors • Execute Setting Menus ZR-RX40 User’s Manual...
  • Page 85 • Both: Operates when a searched alarm is generated or canceled. • H: Operates when a searched alarm is generated. • L: Operates when a searched alarm is canceled. Level: Sets the level to be searched for analog CH and pulse. Setting Menus ZR-RX40 User’s Manual...
  • Page 86 The NAVI menu can be displayed in three modes, Free Running, Recording, and Replay. Operation Description Open Press the [NAVI] key to open the NAVI menu. Close Press the [NAVI] key to close the NAVI menu. Browse explanation Explanation is displayed when an enabled key is pressed. Setting Menus ZR-RX40 User’s Manual...
  • Page 87 Left/right key can be used to change the zone division. Replaying SEARCH Left/right key can be used to perform search. Left: Searches past side Right: Searches future side ZONE Left/right key can be used to change the zone division. Setting Menus ZR-RX40 User’s Manual...
  • Page 88 To cancel key lock by password A password can be set to ZR-RX40 to cancel the key lock. (No password is set at factory default.) <Operation flow> Set the password. Press the , and [ENTER] keys at the same time to display the password setting screen shown below.
  • Page 89: Web Server Function

    • http ....Protocol to access the server. HTTP (Hyper Text Transfer Protocol) • IP address..Type in the IP address of the ZR-RX40 to monitor. • Index.html ..File name. This is fixed to Index.html. Procedure Open the Web browser.
  • Page 90 Zoom........Enlarges only the LCD screen of ZR-RX40. Digital ........Displays the ZR-RX40 measured value digitally. Download of device file ..Allows data captured with ZR-RX40 to be downloaded to your PC via FTP. OMRON Web site ....Accesses to our Web site.
  • Page 91 Remote key operation To operate ZR-RX40 from a remote location, click the corresponding ZR-RX40's panel keys on the screen. OMRON ZR-RX40 OMRON web site WEB Server Function ZR-RX40 User’s Manual...
  • Page 92 Press this key to switch among MONITOR, SPAN, POSITION, and TRACE. ← → ↑ ↓......... Cursor keys Screen update speed..... Specifies the speed in which the screen is updated. Available update speeds are 2, 5, and 10 seconds. WEB Server Function ZR-RX40 User’s Manual...
  • Page 93 Screen update speed..... Specifies the speed in which the screen is updated. Available update speeds are 2, 5, and 10 seconds. Download of device file Allows memory data from ZR-RX40 and data in USB memory device to be downloaded to your PC. WEB Server Function ZR-RX40 User’s Manual...
  • Page 94 MEMO WEB Server Function ZR-RX40 User’s Manual...
  • Page 95: Specifications

    SPECIFICATIONS Standard Specifications Accessory/Option Specifications External Dimensions...
  • Page 96: Standard Specifications

    ± (0.05 % of rdg +1.0 °C) 0 ≤ TS ≤ 1300 ± (0.1 % of rdg +1.0 °C) 0 ≤ TS ≤ 2315 ± (0.1 % of rdg +1.5 °C) ± 0.5 °C Reference contact compensation accuracy Standard Specifications ZR-RX40 User’s Manual...
  • Page 97 Level judgment, window judgment, logic pattern judgment, pulse judgment Clock accuracy (*5) ± 0.002 % (approx. 50 sec/month) 0 to 45 °C, 5 to 85 % RH (15 to 40 °C when the battery is used) Operating environment Standard Specifications ZR-RX40 User’s Manual...
  • Page 98: Main Functions

    2 can be set simultaneously operations Method Realtime and between cursors specified (during data replay) (*2) Data Search Function Search the captured data for the required number of points Search type Channel, Pulse, Logic, Level, Alarm search Standard Specifications ZR-RX40 User’s Manual...
  • Page 99 Data transfer to the PC (realtime memory) functions PC control of the ZR-RX40 Ethernet Web server function (Operation of ZR-RX40, Displays ZR-RX40's functions screen image on browser) FTP server function (Transfers and deletes measured data from internal memory via network)
  • Page 100: Accessory/Option Specifications

    Don't start up the other application software, as WaveInspire works. And don't work other several operation. (For example, screen saver, virus scan program, copying or moving files, searching file etc.) Accessory/Option Specifications ZR-RX40 User’s Manual...
  • Page 101: Battery Pack Zr-Xrb1

    When the battery is running low, measured data is saved and the file is closed automatically. For ZR-RX40, when two battery packs are mounted. When capturing to internal memory at a sampling speed of 1 sec, 20 channel terminals, using new battery packs at +25 °C environment.
  • Page 102: External Dimensions

    External Dimensions (Unit: mm) ZR-RX40 mounting screw holes Two, M4, depth 6 23.5 External Dimensions ZR-RX40 User’s Manual...
  • Page 103: Appendix

    APPENDIX Error Messages and Countermeasures Index Revision History...
  • Page 104: Error Messages And Countermeasures

    Invalid trigger stop settings. When logging is started with the trig- Please set the stop trigger correctly. Press [Enter] key and change the set- ger setting not normally set,it is dis- tings. played.(STOP) Error Messages and Countermeasures ZR-RX40 User’s Manual...
  • Page 105: List Of File Error Messages

    "Disk is full" is usually displayed. cause. Please increase the capacity of the • Disk is full. disk as it erases an unnecessary • Neither the file nor the directory file. are created any further. Error Messages and Countermeasures ZR-RX40 User’s Manual...
  • Page 106 MEMO Error Messages and Countermeasures ZR-RX40 User’s Manual...
  • Page 107: Index

    OTHR settings DISPLAY key Dual View Replaying Part Names and Functions ENTER key Interface EU (Scaling) PC Software External Performing Measurement, Dimensions Precautions to Observe Power On Start Pulse 57, 60, 68 FAST FORWARD key Features Index ZR-RX40 User’s Manual...
  • Page 108 Standad Specifications 102, 103 Standard Specifications Standard Specifications 102, 103 Start side source settings START/STOP key Statistical calculation Temp. Unit 72, 73 Text input TIME/DIV key TRIG settings Trigger level settings USB Drive Mode User USER settings Index ZR-RX40 User’s Manual...
  • Page 109 Index ZR-RX40 User’s Manual...
  • Page 110: Revision History

    Original production August 2007 Rear view added in External Dimensions and minor corrections WEB1 October 2007 Accessories changed. October 2009 Function addition according to software upgrade (Ver 2.14) December 2009 Backup Setting added in DATA settings. Revision History ZR-RX40 User’s Manual...
  • Page 111 IL 60173-5302 U.S.A. Carl-Benz-Str. 4, D-71154 Nufringen, Germany Tel: (1) 847-843-7900/Fax: (1) 847-843-7787 Tel: (49) 7032-811-0/Fax: (49) 7032-811-199 © OMRON Corporation 2009 All Rights Reserved. OMRON (CHINA) CO., LTD. OMRON ASIA PACIFIC PTE. LTD. In the interest of product improvement, Room 2211, Bank of China Tower, No.