GE MAC 5000 Operator's Manual page 170

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Defining the System Setup: Define the ECG Setup
Item
Suppress
Select Yes to prevent the Abnormal ECG and
ABNORMAL and
Borderline ECG 12SL analysis statements from
BORDERLINE
appearing on printed, stored, and transmitted ECG
statements
reports.
Pacemaker pulse
Select Yes to detect very small pacemaker pulses.
enhancer
However, when Pacemaker pulse enhancer is on, the
system should NOT be close to equipment emitting
high frequency radiation. High frequency radiation can
interfere with pacemaker pulse detection.
Preview before
Select Yes to allow the user to always preview a
analysis
recorded ECG before the system analyzes the data.
Storage format
Choose the data compression format of the ECGs
stored on a MUSE CV system:
Select 500 Hz (MUSE Network) to transmit ECGs
to a MUSE CV system using MUSE software
versions 004A or later.
á
The Storage format item does not display if the
MAC 5000 has the ACI-TIPI option installed. With
ACI-TIPI, you can only store 500 Hz ECGs.
14-18
Description
Item
Auto ECG
Choose the ECGs you want your system to
storage
automatically store to a disk:
Select All ECGs to automatically store a recorded
ECG.
Select No ECGs to disable automatic storage of a
recorded ECG.
Select Only ABNORMAL ECGs to automatically
store a recorded ECG that the 12SL analysis
program has classified as abnormal.
Auto ECG
Choose the ECGs you want your system to
transmission
automatically transmit:
Select All ECGS to automatically transmit a
recorded ECG.
Select No ECGS to disable automatic transmission
of a recorded ECG.
Select Only ABNORMAL ECGs to automatically
transmit a recorded ECG that the 12SL analysis
program has classified as abnormal.
Enable ACI-TIPI
Select No to disable the ACI-TIPI function.
Select Yes to enable the ACI-TIPI function.
á
This is an option and might not be installed on your
unit.
MAC 5000 System — 2000657-057, Revision B
Description

Advertisement

Table of Contents
loading

Table of Contents