Semi-Detachable Kerrison Rongeur - Washing, Decontamination & Sterilizing instructions – INS009 iss 004

Intended Use and features

The Semi-Detachable Kerrison Rongeur is indicated for use by suitably trained, medical professionals for cutting or biting bone
during surgery. The Semi-Detachable Kerrison Rongeur differs from other Kerrison Rongeurs in that it has the ability to be
partially opened which provides the advantage of efficient cleaning and decontamination between uses and prior to sterilisation
of the whole instrument compared to a non-detachable version.

Product characteristics

The Semi-Detachable Kerrison Rongeur is available with bite sizes between 1 and 5mm in 45 degree or 90 degree upward
cutting versions. They have a working length of 180mm, 200mm or 210mm and have either a reflection reducing or black finish.
Item #
PH654950 to PH654954
PH654955 to PH654959
PH654960 to PH654964
PH654965 to PH654969
PH654970 to PH654974
PH654975 to PH654979
PH654980 to PH654984
PH654985 to PH654989
PH654990 to PH654994
PH654995 to PH654999
PH655000 to PH655004
PH655005 to PH655009
Y
Correct bite size selection based on the surgical discipline is of utmost importance to prevent overloading which could
cause damage to the instrument.
Y
Device proper functionality including assurance of full assembly shall be verified prior to each use.
The lifetime of the Semi-Detachable Kerrison Rongeur is determined by proper functionality and careful care during use. End of
lifetime is determined by wear and damage due to use. Improper functioning, damaged or excessively worn Semi-Detachable
Kerrison Rongeur should not be used.
Y
Contra-Indications
The Semi-Detachable Kerrison Rongeur is contra-indicated for use in contact with the heart, central circulatory system and
central nervous system.
INS009 issue 004 Semi Detachable Kerrison Rongeur - Washing, Decontamination & Sterilizing
instructions
Semi-Detachable Kerrison Rongeur Upper Cut / 180mm - 45 Degree 1mm to 5 mm
Semi-Detachable Kerrison Rongeur Upper Cut / 180mm - 90 Degree 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 200mm - 45 Degree 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 200mm- 90 Degree 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 210mm - 45 Degree 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 210mm - 90 Degree 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 180mm - 45 Degree / Black / 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 180mm - 90 Degree / Black / 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 200mm - 45 Degree / Black / 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 210mm - 45 Degree / Black / 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 200mm - 90 Degree / Black / 1mm to 5mm
Semi-Detachable Kerrison Rongeur Upper Cut / 210mm - 90 Degree / Black / 1mm to 5mm
Page 1 of 4
P
STERIS Ireland Limited.
IDA Business and Technology Park,
Tullamore, County Offaly,
R35 X865, Ireland
Product Description
Effective 2020-12-14
C
C

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH654950 and is the answer not in the manual?

Questions and answers

Summary of Contents for Steris PH654950

  • Page 1 M STERIS IMS Ltd. 14 Pindar Road, Hoddesdon, Hertfordshire, STERIS Ireland Limited. EN11 0BZ, England IDA Business and Technology Park, Telephone No. +44 (0)3452 414818 Tullamore, County Offaly, E-Mail: instrumentinfo@steris.com R35 X865, Ireland Web: www.sterisims-instruments.com Semi-Detachable Kerrison Rongeur - Washing, Decontamination & Sterilizing instructions – INS009 iss 004...
  • Page 2 Cleaning & User Instructions This device is supplied NON-STERILE: To be sterilized before use. The device can be damaged due to inappropriate cleaning, disinfection and/or sterilization. Physical removal of gross debris and bioburden, through soaking and wiping of instrumentation with water should be initiated immediately after the procedure, at the point of use.
  • Page 3 Packaging materials used should conform to EN ISO 11607 or the relevant parts of EN 868-7 For further information, related to the use of these instruments, please contact your STERIS Instrument Management Services Limited representative. These instruments must be disassembled in order to be washed, cleaned and decontaminated effectively.
  • Page 4 Serious Incidents Serious incidents that have occurred in relation to this medical device should be reported to a STERIS Instrument Management Services representative and competent authority in the country where the incident occurred. Disposal after use The device is made from metal. The device itself does not require special disposal instructions. The device should be cleaned and disinfected to remove any possibility of infection.