Performance Verification

Pulse Oximetry Functional Tests

Introduction
26
5. Press the ON/STANDBY button to turn the monitor on and verify that the
monitor is operating.
6. The monitor should stabilize on the subject's physiological signal in about 15
to 30 seconds. Verify that the oxygen saturation and pulse rate values are
reasonable for the subject.
These tests utilize the pulse oximetry functional tester (Nellcor model
SRC-MAX) to verify the performance of the N-595 monitor. See Figure 3.
All of these tests should be done in sequence.
1 DOC-10 Cable Connector
2 RED LED Drive Indicator
3 Not Used For N-595
4 Not Used For N-595
5 Battery Low Indicator
Figure 3: SRC-MAX O
The SRC-MAX functional tester allows qualified technicians to functionally test
Nellcor O
M
technology-based pulse oximeters and OEM O
XI
AX
technology-based monitors. The technician must perform the test setup procedure
6 % Modulation Select Button
7 % SpO
Select Button
2
8 Light Level Selection Button
9 Pulse Rate Selection Button
10 INFRARED LED Drive Indicator
M
Oximetry Tester
XI
AX
M
XI
AX