Performance Verification
34
•
Earth Leakage Current on page 102 for test values.
•
Enclosure Leakage Current on page 102 for test values.
•
Patient Applied Risk Current on page 102 for test values.
•
Patient Applied Risk Current on page 102 for test values.
Note:
Patient Applied Risk Current and Patient Isolation Risk Current: The
leakage test lead from the test equipment must be connected to the N-595
SpO
Sensor Port through the DOC-10 pulse oximetry cable using a male
2
9-pin "D" type connector that has all pins shorted together. During these
tests the monitor will display "EEE 10," after the "Nellcor" screen.