Performance Verification
34
Earth Leakage Current on page 102 for test values.
Enclosure Leakage Current on page 102 for test values.
Patient Applied Risk Current on page 102 for test values.
Patient Applied Risk Current on page 102 for test values.
Note:
Patient Applied Risk Current and Patient Isolation Risk Current: The
leakage test lead from the test equipment must be connected to the N-595
SpO
Sensor Port through the DOC-10 pulse oximetry cable using a male
2
9-pin "D" type connector that has all pins shorted together. During these
tests the monitor will display "EEE 10," after the "Nellcor" screen.