Specifications
96

Table 14: Electromagnetic Immunity

The N-595 is suitable for use in the specified electromagnetic environment. The
customer and/or user of the N-595 should assure that it is used in an electromagnetic
environment as described below.
Immunity
IEC 60601-1-2
Test
Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electric fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV
differential
mode
IEC 61000-4-5
±2 kV common
mode
Note: U
is the AC mains voltage prior to application of the test level.
T
Electromagnetic
Compliance
Environment
Level
Guidance
±6 kV contact
Floor should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
±8 kV air
relative humidity should
be at least 30 %.
±2 kV for power
Mains power quality
supply lines
should be that of a typical
commercial and/or
hospital environment
±1 kV for input/
output lines
±1 kV
Mains power quality
differential
should be that of a typical
mode
commercial and/or
hospital environment
±2 kV common
mode