Safety Tests
Ground Integrity
N-595
Cables and
Maximum
Sensors
Length
O
M
3 feet (0.9 m)
XI
AX
Durasensor
sensor
DS-100A
O
M
OxiCliq
OC-3 cable
XI
AX
sensors:
3 feet (0.9 m)
P
N
I
A
O
M
4 feet (1.2 m)
XI
AX
Dura-Y sensors:
D-YS
D-YSE
D-YSPD
100 milliohms or less
Table 17: Cables
Complies With
•
RF emissions, CISPR 11,
Class B/Group 1
•
Harmonic emissions, IEC 61000-3-2
•
Voltage fluctuations/flicker emission,
IEC 61000-3-3
•
Electrostatic discharge (ESD),
IEC 61000-4-2
•
Electric fast transient/burst,
IEC 61000-4-4
•
Surge, IEC 61000-4-5
•
Conducted RF IEC 61000-4-6
•
Radiated RF, IEC 61000-4-3
•
RF emissions, CISPR 11,
Class B/Group 1
•
Harmonic emissions, IEC 61000-3-2
•
Voltage fluctuations/flicker emission,
IEC 61000-3-3
•
Electrostatic discharge (ESD),
IEC 61000-4-2
•
Electric fast transient/burst,
IEC 61000-4-4
•
Surge, IEC 61000-4-5
•
Conducted RF IEC 61000-4-6
•
Radiated RF, IEC 61000-4-3
Specifications
101