Setting The Resolution And Polarity - Sony Magnescale LH51 Instruction Manual

Display unit
Table of Contents

Advertisement

Available languages

Available languages

5-1-1. Setting the resolution and polarity

• Select the resolution and polarity setting mode by pressing the X-axis selector key when the
• The minus "–" display indicates that the polarity is inverted.
• The resolution and polarity can be set and changed with the
• The resolution is automatically set for Magnescale, GA or Digiruler axes, so only set the polarity
Example : Input to the first axis (Magnescale axis)
X
0
• When connecting other scales using the DZ51, the resolution can be selected from 0.0001 mm,
display is as shown in 5-1.
in these cases.
For Magnescale axes : the scale resolution is automatically set to 0.0005 mm
For GA axes
: the scale resolution is automatically set to 0.005 mm
For Digiruler axes
: the scale resolution is automatically set to 0.01 mm
Operating Procedure
Select the input to the first axis.
The scale resolution is automatically set to
0.0005 mm.
Press the
key to increase the value.
0.0005 mm, 0.001 mm, 0.005 mm, 0.01 mm, 0.025 mm, 0.05 mm and 0.1 mm.
Set the resolution to match that of the connected scale.
Display
.0001
.0005
.001
.005
.01
.025
.05
.1
Resolution
0.0001 mm
0.0005 mm
0.001 mm
0.005 mm
0.01
mm
0.025 mm
0.05
mm
0.1
mm
key or
key.
Display
13