Nibp Setup Screen; Mode; Target - GE Corometrics 250cx Series Operator's Manual

Hide thumbs Also See for Corometrics 250cx Series:
Table of Contents

Advertisement

NIBP Setup Screen

kPa mode
High
Low
6.7-20.0
9.3-32.0
9.3-17.3
4.0-16.0
9.3-20.0
130/ 85
4.0-16.0
100-250
35-120
mmHg mode
High
Low
50-150
70-240
70-130
30-120
70-150
30-120
100-250
35-120

Mode

Target

Revision C
Maternal Non-Invasive Blood Pressure Monitoring: NIBP Setup Screen
Select the NIBP softkey to access the NIBP Setup screen.
FECG
US
165 172
NIBP Setup
NIBP
11:41
MAP (107)
NIBP Done Vol:
03:15
Alarms
High
Systolic:
Diastolic:
MAP:
MHR/P:
Alarm Volume:
This field alternates between the manual and automatic monitoring modes for
maternal blood pressure. For auto mode, this field also sets the interval time, in
minutes, between automatic blood pressure determinations. This interval time is
measured from beginning to beginning of determinations. (The monitor is factory-
set with the optional 1-minute interval time enabled. For information on disabling
the 1-minute interval, refer to the "250/250cx Series Monitor Service Manual".)
NOTE: As soon as the auto mode is selected on the setup screen, the countdown
timer begins to decrement. The first automatic determination begins after
expiration of one complete interval time period.
This option lets you choose the initial pressure for cuff inflation. If the NIBP is taken
while previous determination is still displayed, and within 16 minutes of the
previous determination, the initial target pressures for subsequent determinations are
based upon the systolic values of previous determination. The default initial target
pressure is 135 mmHg (18.0 kPa). Adjust the pressure between 100 to 250 mmHg
(13.3 to 33.3 kPa) in increments of 5 mmHg (0.7 kPa).
NOTE: Selecting a target pressure will clear old NIBP values in the vital signs area
and cancel any determination in progress.
250cx Series Maternal/Fetal Monitor
2036946-001
TOCO
30
Manual
Mode:
Target:
135
5
Low
160
90
mmHg
90
50
mmHg
140
50
mmHg
120
50
bpm
5
Exit
Manual
Auto 1 min
Auto 2 min
Auto 3 min
100-250,
Auto 4 min
5 mmHg
Auto 5 min
increments
Auto 10 min
Auto 15 min
0
Auto 20 min
1
1
Auto 30 min
2
2
Auto 40 min
3
3
Auto 45 min
4
4
Auto 60 min
5
5
Auto 90 min
6
6
Auto 120 min
7
7
8
8
9
9
Maternal NIBP Setup Screen