Printing the Maternal Vital Signs History Screen

NOTE: Once the print function is
activated, you may exit the history
screen; it need not remain displayed
during printing.

Printing the Entire Vital Signs History

Printing a Page of the Vital Signs History

Stopping the Printing of Maternal Vital Signs History

Revision C
Maternal Vital Signs History: Using the Maternal Vital Signs History Screen
When the screen is called up for display, the previous minute in time is listed at the
far right of the screen; the preceding values to be displayed are counted backwards
from the previous minute in time, based on the intervals you have selected.
Example: The HX Interval is set at 5 minutes and the maternal Vital Signs History
screen is displayed when the current time is 13:57. When the screen is displayed
there will be five columns of data for the following times:
13:36
If the screen remains displayed for 1 minute the columns of data change to:
13:37
You can select all or a portion of the maternal vital signs history for printing on the
strip chart paper. The information is printed in the upper portion of the top (or left)
grid of the strip chart paper at a high speed mode. If the recorder is on, all other
trending is interrupted while the maternal vital signs history is printed. If the
recorder is in the maternal-only mode, the recorder interrupts any printing of data to
print the vital signs history; any pre-empted data is printed at the end. If the recorder
is off, it will turn on to print the vital signs history, then turn off again.
Select the PrintAll softkey to print the entire vital signs history. An audible tone
provides confirmation.
View
Use the
selected page. An audible tone provides confirmation.
Printing automatically stops if:
you open the recorder door;
strip chart paper runs out;
you press the
you change the recorder mode.
CAUTION
Do not turn monitor off during printing or your data will be lost.
250cx Series Maternal/Fetal Monitor
13:41
13:46 13:51
13:56
13:42
13:47
13:52
13:57
softkey to display a page; then select the Print softkey to print the
Test
button;