Functionality with a QS System
Paper Versus Electronic Strip Charts
Revision C
Recorder Modes: Functionality with a QS System
CAUTION
DATA STORAGE—Stored maternal vital signs history data is
erased when you turn the monitor off. Therefore, for intermittent
monitoring, it is recommended that you leave the monitor on, but
turn the recorder off. Refer to Chapter 12, "Maternal Vital Signs
History" for more information.
From
To
Off
On
Off
Maternal-Only
On
Maternal-Only
On
Off
Maternal-Only
On
Maternal-Only
Off
1
A confirmation tone sounds to indicate the recorder has been turned off.
Recorder Mode Audiovisual Indicator Status
Recorder Mode
On
Maternal-Only
Off
Users of a Quantitative Sentinel (QS) System (Software Version 4.0.3.0 or earlier)
should be aware of the following items when using the monitor's maternal-only
mode:
As described earlier, the monitor's maternal-only mode acts as a "paper saver"
turning the strip chart recorder on and off as needed. However, the QS System
overrides the maternal-only mode by storing the entire patient record. In other
words, the electronic strip chart is retained as if the monitor's recorder were left on
250cx Series Maternal/Fetal Monitor
2036946-001
Changing Recorder Modes
Button
Briefly press once; or press and hold 3 seconds.
Briefly press twice.
Briefly press once.
Press and hold 3 seconds.
Briefly press once.
Press and hold 3 seconds.
Record Indicator
continuously illuminated
lights three short flashes every
5 seconds
Off
1
1
Audio Indicator
Off
11-5