HR/PR Trace

Volume

Alarms

Alarm Volume

MECG Lead

Revision C
Maternal Heart/Pulse Rate Monitoring: MHR/P Setup Screen
This field enables or disables the printing of the MHR/P trace on the strip chart
paper.
On: The MHR/P trend is printed in grey annotated with
—whichever parameter is selected in the MHR/P Source field. MHR/P
P
data from NIBP is not trended since blood pressure determinations are static
measurements.
Off: The MHR/P trend is not printed.
This field sets the volume of the "beep" sounded with each detected valid
heartbeat—for MECG and MSpO
These fields adjust the high and low alarm limits for MHR/P— in increments of 5
bpm. The selectable values are shown in the MHR/P Setup screen. The factory
defaults are listed in Appendix A, "Factory Defaults".
This field controls alarm volume for all maternal alarms.
This field selects the ECG lead configuration. The lead can also be selected from the
MECG Lead Softkey on the normal operating screen.
Lead I refers to the potential between the left arm and the right arm.
Lead II refers to the potential between the right arm and left leg.
Lead III refers to the potential between the left arm and the left leg.
250cx Series Maternal/Fetal Monitor