Mecg Pacer - GE Corometrics 250cx Series Operator's Manual

Hide thumbs Also See for Corometrics 250cx Series:
Table of Contents

Advertisement

WHITE (RA)
RED (R)

MECG Pacer

The bracket situated to the left of the ECG waveform denotes 1 mv. If the ECG
waveform size is set to Auto, the bracket will auto-adjust to maximize the QRS
display, depending on signal amplitude.
7-6
Maternal Heart/Pulse Rate Monitoring: MHR/P Setup Screen
The following figure illustrates which electrodes reference the ECG lead obtained.
BLACK (LA)
YELLOW (L)
RED (LL)
GREEN (F)
NOTE: AHA label is bolded; IEC label is italicized.
This field enables/disables pacemaker pulse rejection circuitry.
Off: Use this setting for a patient without a pacemaker. All ECG events are
monitored; all complexes, including pacer spikes may be displayed
included in the heart rate calculation.
On: Use this setting for a patient who has a pacemaker. The monitor rejects the
pacer spike from the heart rate calculation and replaces the actual pacer spike
with a pacer event mark; in addition the letter
waveform speed. Following is an example of an MECG waveform with the
MECG pacer On.
MECG
09:21:41
Frozen
Print
Freeze
MECG Waveform with Pacer Enabled
1
If the MECG waveform is enabled for display
250cx Series Maternal/Fetal Monitor
2036946-001
WHITE (RA)
RED (R)
Alarms
Setup
BLACK (LA)
YELLOW (L)
RED (LL)
GREEN (F)
MECG Lead Selection Guide
1
and may be
P
is displayed prior to the
P 25 mm/s
II Auto
09:22:06
VSHX
Revision C
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Corometrics 256cxCorometrics 259cx

Table of Contents