Cleaning; Basic Cleaning - BRAND seripettor Operating Manual

Bottle top dispenser
Hide thumbs Also See for seripettor:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 37

8 Cleaning

Components filled with reagent
The cylinder, valves, telescopic filling tube and discharge tube contain reagent!
> Never remove the discharge tube while the dispensing cylinder is filled.
> Point the valves and tube openings away from your body.
> Wear protective clothing, eye protection and protective gloves!
The instrument must be cleaned in the following situations to assure correct operation:
immediately when the piston is difficult to
move
prior to a reagent change
prior to long term storage
prior to dismantling the instrument
prior to changing the valve

8.1 Basic cleaning

1. Completely emptying the instrument
a.
Screw the instrument onto an empty bottle and empty it completely by dispensing. If the instru-
ment is equipped with a recirculation valve, it must be emptied in both the 'dispense' and 'recir-
culate' settings.
2. Rinsing the instrument
a.
Screw the instrument onto a bottle filled with a suitable cleaning agent (e.g. deionized water)
and rinse the instrument several times by completely filling and emptying it.
997468
WARNING
prior to autoclaving
regularly when using liquids which form de-
posits (e.g., crystallizing liquids)
regularly when liquids accumulate in the
screw cap.
8 Cleaning
55
Operating manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Seripettor pro

Table of Contents