Quality Control (QC)

A quality control program should be performed on a regular basis to verify the MyCare Insite Analyser
is working properly and providing reliable results. Data integrity can only be assured when controls and
Good Clinical Laboratory Practice are used routinely. The frequency of performing QC differs from
laboratory to laboratory. All quality control requirements and testing should be performed in
accordance with local, state and/or federal regulations or accreditation requirements.

QC Materials

Quality control material is defined in the MyCare Insite Test package inserts.
To run QC:
Prepare Cuvette and Reagent Cap - allow at least 10 minutes to come to room temperature
-
Place the RFID card provided in the test box in the RFID card well on top of the Analyser
-
Prepare the control according to the instructions in the control package insert
-
Enter the control level and control lot number instead of the patient data
-
Add exactly the volume of control specified in the test package insert to the Cuvette
-
Prepare the test according to the Measuring QC instructions in the package insert
-
Insert the cartridge into the Analyser and start the test measurment by closing the door
-
At the end of the measurement:
-
The door will open; remove cartridge and close door
The result is displayed on the Touch Screen
-
Record the result according to your laboratory's quality guidelines
-
The result will be saved in the Touch Screen's memory – just like a patient's result
-
If desired, export the result(s) to an (optional) external computer, print the result with an
-
(optional) external printer or send it via email (see page 30)
Verify that the result is within the limits specified for the MyCare Insite control material
-
(according to the control package insert)
Be sure to follow the detailed instructions for the test that are provided in the operation sections (page
28) and comply with the information provided in the package insert at MyCareInsite.com.

Handling QC Control Materials

Please refer to the package insert provided with each control kit for detailed instructions
on storage and handling of control materials.
Follow the instructions in the "Testing a Patient Sample" sections (starting on page 28) to properly
perform a control test. The measured values must be within the range of target values specified in the
control package insert. If the control results fall within the specified range, the testing of patient
samples may begin.
PI MCI-EUR Rev 05