Indicator Lights

MyCare Insite Analyser front

How to Handle the MyCare Insite Analyser Laboratory Photometer

The Analyser is operated solely by means of the Touch Screen. All the basic operating steps are
displayed as symbols in Table 1 (page 2) and Table 2 (page 3).
To activate a symbol tap it with a finger.
In order to perform a test, the RFID card enclosed in the test box must first be placed on the Analyser. This
card contains all the data needed to perform the tests. No analysis can be started without the RFID test
card.
The door opens automatically when a test is initiated by pressing the
requested data on the Touch Screen and inserting the test cartridge (Cuvette with Reagent Cap) into the
Chamber, close the door. The test starts automatically. After the test is completed, the door opens
automatically, remove the test cartridge, and close the door again.
The door prevents ambient light, dust, dirt, and humidity from entering the Analyser during the testing
process, and when the Analyser is not in use.
Please make sure the door is closed whenever the Analyser is not in use.
PI MCI-EUR Rev 05
LEDs on the MyCare Insite Analyser
This green LED indicates the Analyser status
LEDs
This blue LED indicates Touch Screen connection status
This blue LED indicates Bluetooth connection status
Note: Use of the USB cable connection is recommended to keep
the Touch Screen powered.
Flashing: the Analyser is warming up and testing cannot be
Note
started yet. (
: warm-up takes approximately 10 min.)
On (green) : the Analyser is warmed up and ready for use
Off: the Analyser is not plugged in
On (blue) : connection with Touch Screen is established
Off: the Touch Screen is not connected
On (blue) : connection with Touch Screen is established
Flashing: Analyser Bluetooth is activated but no Touch Screen
is connected via Bluetooth
Off: Bluetooth is switched off
button. After entering all the